Adjusting the tilt angle
Using knob (1), you can position the surgical microscope in a range from
+90° to -90° (+ in the direction of the surgeon and - in the opposite direc-
tion). The +90° setting is ideal for surgery on patients in a seated position
or lying on their side.
Caution:
Do not tilt the main microscope beyond + / -90°, as this could damage the
microscope cable or the light guide.
Turn knob (1) until the surgical microscope is in the viewing position
required.
After the viewing angle has been set, the surgical microscope remains in
this position. The gear drive is self-locking.
Warning!
To prevent the integrated assistant's microscope from moving down-
ward of its own accord when the main microscope is being tilted, the
integrated assistant's microscope must be adjusted and locked in po-
sition using screw (2) before surgery.
Check that the integrated assistant's microscope is firmly seated.
Do not tilt the main microscope beyond + / -90°, as this could damage
the microscope cable or the light guide.
Using locking screw (2), secure the integrated assistant's microscope
in the selected working position (on the left or right of the main micro-
scope).
Turn knob (1) until the surgical microscope is in the viewing position
required.
After the viewing angle has been set, the surgical microscope remains in
this position. The gear drive is self-locking.