Technical data
170 Preparations
G-30-1682-en OPMI® Lumera® T Issue 4.0
Printed on 02. 02. 2009
Settings on the control and display panel
Setting up the suspension system
• Turn on the suspension system at its power switch.
• Successively select the following functions on the suspension system:
– lamp brightness,
– motor speeds for zoom,
– focus and
– X-Y coupling.
• Set the lamp brightness as follows:
Slowly increase the brightness until the necessary and still admissible
level is reached.
– Halogen: adjustment range: 0.5 ... 10
– Xenon: adjustment range: 0.5 ... 10
Warning!
Too much light intensity (brightness setting is too high) or excessive radi-
ation exposure times may lead to retinal injury in the patient's eye.
• Adjust the illumination intensity as required for the selected type of il-
lumination and the radiation exposure time. You will find the values
recommended by Carl Zeiss in the table "Maximum radiation expo-
sure times" on page 29.
• Set the values required for
– motor speeds of the functions zoom,
– focus and
– X-Y coupling.
Motor speed
Adjustment range: 1...10