Technical data

170 Preparations
G-30-1682-en OPMI® Lumera® T Issue 4.0
Printed on 02. 02. 2009
Settings on the control and display panel
Setting up the suspension system
Turn on the suspension system at its power switch.
Successively select the following functions on the suspension system:
lamp brightness,
motor speeds for zoom,
focus and
X-Y coupling.
Set the lamp brightness as follows:
Slowly increase the brightness until the necessary and still admissible
level is reached.
Halogen: adjustment range: 0.5 ... 10
Xenon: adjustment range: 0.5 ... 10
Warning!
Too much light intensity (brightness setting is too high) or excessive radi-
ation exposure times may lead to retinal injury in the patient's eye.
Adjust the illumination intensity as required for the selected type of il-
lumination and the radiation exposure time. You will find the values
recommended by Carl Zeiss in the table "Maximum radiation expo-
sure times" on page 29.
Set the values required for
motor speeds of the functions zoom,
focus and
X-Y coupling.
Motor speed
Adjustment range: 1...10
Level 1 corresponds to the lowest, level 10 to the highest motor speed.