Technical data

154 Preparations
G-30-1682-en OPMI® Lumera® T Issue 4.0
Printed on 02. 02. 2009
Connections
Connecting the surgical microscope
Turn locking cap (1) by a quarter turn to the right or left and pull up cov-
er (2).
Plug microscope connector (3) into connector (4) and tighten the se-
curing screws on the microscope connector.
Press the microscope cable into cable clip (5). Install the cable in such
a way that it is neither stretched nor kinked when the microscope is
turned or tilted.
Press down cover (2) as far as it will go and lock it with cap (1).
Connecting the light guide
Warning!
Risk of phototoxic injury to the patient's eye.
Only operate the VISULUX
TM
fiber slit lamp with the Superlux Eye light
source if the light guide sleeve contains a UV filter. You can recognize an
integrated UV filter by the silver-colored
light guide connector of the VISU-
LUX
TM
fiber slit lamp.
Insert the end of the light guide into light guide connector (7) of the mi-
croscope as far as it will go, and press the light guide into cable
clip (6). For correct mounting, please see the label provided under
light guide connector (7).
Note:
Make sure that the light guide is not stretched or bent when the micro-
scope is turned or tilted.