Technical data
150 Preparations
G-30-1682-en OPMI® Lumera® T Issue 4.0
Printed on 02. 02. 2009
Warning!
• Before every use and after re-equipping the system, make sure that
the two mounted tubes (3 and 13) are securely locked in position.
• Make sure that the following components are firmly
tightened:
– securing screws (6 and 8) and
– objective lens (7)
• When attaching any components, take care not to damage the Deep-
View system.
• Loosen locking knob (1).
• Re-adjust the balance of the suspension arm after every change of
equipment.