Technical data

148 Preparations
G-30-1682-en OPMI® Lumera® T Issue 4.0
Printed on 02. 02. 2009
Mounting the tube, the eyepieces and the objective lens
Note:
The main microscope comes with Invertertube (3), the integrated as-
sistant's microscope with assistant's Invertertube (13). Alternatively, the
binocular coobservation tube (10) and the 45° inclined tube (11) can be
attached.
Bring the suspension arm in a position convenient for you and firmly
tighten locking knob (1).
Give securing screw (6) a few turns to loosen it.
Remove dust cover (2) and store it in a safe place.
Place Invertertube (3) or binocular tube (10) on the microscope and
firmly
tighten securing screw (6).
You can install further accessories between Invertertube (3) or binoc-
ular tube (10) and the microscope body. Lock these modules in posi-
tion in the same way using securing screw (8).
Place assistant's Invertertube (13) or 45° inclined tube (11) on the in-
tegrated assistant's microscope and firmly
tighten securing screw (8).
You can install further accessories between assistant's Invertertube
(13) or 45° inclined tube (11) and the microscope body. Lock these
modules in position in the same way using securing screw (8).
Insert 10x widefield eyepieces (5) as far as they will go in mounts (4)
intended for them. The magnetic coupling reliably secures them in po-
sition.
Insert 10x widefield eyepieces (12) as far as they will go in mounts (9)
intended for them. The magnetic coupling reliably secures them in po-
sition.
Note:
If you wish to use documentation equipment, we can supply an eyepiece
with a reticle to aid focusing. The retrofitting of a reticle to an eyepiece can
only be performed in the factory or by our service staff. Always install the
eyepiece with the reticle on the same side of the binocular tube where the
documentation equipment is located.
Screw objective lens (7) into the microscope body and tighten it firmly
.