Technical data

G-30-1682-en OPMI® Lumera® T Issue 4.0
Printed on 02. 02. 2009
Description 127
Description of components
The S81 ceiling mount comprises a column, a carrier arm and a suspen-
sion arm.
The suspension arm with the lamp housing and the control unit is
mounted on the carrier arm. The control unit is rotatable through 180° (70°
if the Superlux Eye light source with additional, integrated halogen light
source is used) and contains all electrical supply systems required for the
control of a motorized surgical microscope. The motorized functions can
be activated using a foot control panel.
The suspension arm permits almost effortless positioning of the surgical
microscope. The spring force of the suspension arm can be varied in a
range from 8 to 20 kg, permitting reliable balancing of the microscope
even with heavy accessory equipment. The downward movement of the
suspension arm can be limited as required.