Technical data
108 Description
G-30-1682-en OPMI® Lumera® T Issue 4.0
Printed on 02. 02. 2009
Connection panel
1 Remote socket
for triggering an AUX signal, e.g. to switch on/off an external device
operating at max. 24V/0.5A.
2 Connector for switching component
Possibility of connecting a foot control panel or operating chair with an
appropriate footswitch.
3 Potential equalization connector
4 Indicator window for rated voltage
The voltage shown here must correspond to the rated line voltage pro-
vided on the site of installation. You can adjust the sliding switch using
a suitable tool.
Warning!
Please observe the maximum current consumption of two power outlets
(5) and (6). Only connect medical devices approved by us to outlets (5)
and (6). When using other devices, make sure that safety is guaranteed
regarding admissible ground leakage currents. The admissible limit value
of the ground leakage current present in the suspension system's power
cord must not exceed 500 µA in compliance with EN60601-1/IEC 60601-
1. CSA/US certification in compliance with UL 60601-1 only allows a max-
imum ground leakage current of 300 µA.
5 Power outlet
for medical devices with a current consumption of max. 2 A.
Note:
The current of this power outlet is switched on/off using the S2 power
switch (7).
6 Power outlet socket
for medical devices with a maximum current consumption of 5 A.
7 Power inlet socket
8 Power switch S2
When the suspension system is on, the green indicator light in the
switch is lit.
9 Strain relief device
The strain relief device prevents inadvertent unplugging of the fol-
lowing electrical connections:
– power cable,