3 Technical Data
Parameters Description
Ambient conditions
Operation Storage Transport
Temperature + 10°C ... + 35°C 0°C ... + 35°C
20°C ... + 50°C
Humidity max. 75% at 35°C
< 90% at 30°C
condensation water not
admissible
< 90% to 30°C
condensation water not
admissible
Conditions of use Dust-free environment, absorption-free worktable
Safety
Protection degree I
Protection type IP 20
Excess voltage category II
Pollution degree 2
Radio interference suppression in accordance with EN 55011 (Class B)
Insensitivity to noise in accordance with EN 50082-1
The components for 3D microscopy meet the requirements of the EC directive
89/336/EEC and the EMC legislation of November 9, 1992
Power supply
3D-CONTROL DSA-3
230 VAC 10%, 50/60 Hz
switchable to 115 VAC 10%, 50/60 Hz
Power consumption of the overall
system
30 VA; minimal 7 VA; maximal 200 VA (depending on configuration)
Fuses
Line input
3D-CONTROL DSA-3
115 VAC: T 0,2 A/250 V in accordance with IEC 127
230 VAC: T 0,1 A/250 V in accordance with IEC 127
For MCU II, monitor shutter, monitor and microscope, see the enclosed Technical Data sheets.