Technical data

Image Acquisition
34 Printed 10.2013
If several cameras are operated on a system (e.g. an AxioCam MRm
and an AxioCam IC), the reference images for each camera will be
managed separately.
In the Multidimensional Acquisition function, a function for
fluorescence-channel-specific shading correction is offered under
Extended parameters on the C (Channel) property page.
Activating shading correction can reduce the maximum possible speed
of the live image.
Applying object-specific shading correction
Activate the Automatic check box on the camera property page General.
Note:
Make sure that the shading reference images generated are still having
the desired effect if the images were generated some time ago. They
will need to be generated again in any case if the optical settings of the
microscope have been changed. This applies to the following settings
in particular:
Lamp adjustment
Köhler setting
Optovar
Camera adapter
Camera port
Applying shading correction manually
With non-motorized and non-encoded microscopes the automatic shading
correction cannot be activated. But it is possible to carry out shading
correction using a previously generated standard reference image. This
image is applied irrespective of the objective set.