6HFWLRQ 2SHUDWLRQV2YHUYLHZ
0DQXDO &DOLEUDWLRQ
During this procedure, the printer determines the media type, label length, media and
ribbon sensor settings, and printing method. Media type is determined by sensing either
continuous or non-continuous media as blank labels move through the printer. If non-
continuous media is sensed, label length is also calibrated. If ribbon is sensed, the
thermal transfer print method is configured. If ribbon is not sensed, the direct thermal print
method is configured.
The results of the calibration are stored in the printer’s memory and are retained even if
power is removed. These parameters remain in effect until the next calibration is
performed.
1. Turn printer power Off (O).
2. Place the head-open lever in the OPEN position.
3. Remove the ribbon.
4. Remove approximately six labels from the media roll, enough so only the liner is
threaded under the media sensor.
5. Reload the media.
6. Press and hold PAUSE, FEED, and CANCEL.
7. Turn the power switch On (l).
8. After the calibrate LED goes out, release PAUSE, FEED, and CANCEL.
9. When the printhead LED flashes, reload the ribbon.
10. Ensure the media sensor is properly positioned.
11. Close the printhead.
12. A media and ribbon sensor profile prints.
Important • Perform manual calibration when using preprinted media, when in
Manual Calibration Mode, or if auto calibrate does not work.
Notes • Perform this procedure only once to put the printer into manual calibration.
After that, press MODE three times and FEED once when changing media (a label
prints showing the sensor profile).
If the printer is in Peel-Off Mode, the operator must catch the labels as they peel away
from the liner.
Note • To return to auto calibration, press and hold PAUSE, CANCEL, and MODE
when you turn the printer on.