User manual
– when assembling or disassembling the lamp
housing,
– before opening the lamp housing,
– when changing the high-pressure mercury
burner and remaining parts, such as the heat-
absorbing filter or the collector,
– for service work on the supply unit.
Lamp housing
Leica MacroFluo™ with fluorescence illuminator
may be used only with the Leica lamp housing
106 z for 100-W mercury bulb, supply unit and
stray-light protection.
For assembly work, always pull the
power plug of the supply unit from the socket!
Before inserting the burner, disconnect the power
supply connection. Adjust the burner immediately
after ignition. See also the special guideline.
Leica Microsystems assumes no liability
for damages if bulbs from other manufacturers
are used.
– Minimum distance of lamp housing (hot sur-
faces!) 10cm (4") from flammable objects such
as books, curtains, wallpaper. Fire hazard!
– Never look into the direct beam path! Danger of
blinding!
– Never open the lamp housing when the lamp is
switched on. Disconnect the power plug of the
supply unit from the supply system.
Danger of explosion, UV radiation, danger of
blinding!
– Let the lamp housing cool off for at least 15 min.
Danger of explosion, danger of burns.
– Never cover the venting slots on the lamp hous-
ing. Fire hazard!
– Use stray-light protection.
UV radiation can damage the eyes.
9
Direct contact with eyepieces is a poten-
tial transmission method for bacterial and viral
infections of the eye. Users should be made
aware of the potential risk of infection. The risk
can be kept to a minimum by using personal eye-
pieces for each individual or detachable eyecups.
1.3 Light source: Safety instructions
Protective measures of the manufacturer:
– UV protection screen in front of the object
plane prevents direct view onto the UV beams.
– The design prevents direct UV radiation on the
eyes in case of empty positions of the nose-
piece disc.
– UV filter in the observation beam paths protects
the eyes.
– Stray-light protection on the lamp housing pre-
vents irradiation of the hands.
Warning: UV radiation could damage
the eyes.
For this reason:
– never look at the light spot on the object plane
without UV protection screen.
– do not select a white, heavy reflecting back-
ground for the objective.
– Never open the lamp housing while the lamp is
switched on.
Danger of explosion, UV radiation, danger of
blinding!
Supply unit
Do not connect the unit until the lamps have been
assembled. Check the setting of the power supply
voltage and correct it, if necessary. See also the
special guideline.
Always disconnect the power plug of the supply
unit from the supply system:
Leica MacroFluo – Assembly