® U SE R MA N UA L
C o n t en t s Warranty............................................................................................................... i Introduction........................................................................................................ 1 Getting Started.................................................................................................... 1 Initial Inspection.................................................................................. 1 Battery Installation...............
Dissolved Oxygen (DO).................................................................... 21 Sensor Maintenance........................................................................... 60 Barometer........................................................................................... 31 Sensor Storage.................................................................................... 66 Conductivity....................................................................................... 32 pH.......
Warran t y This page lef t in t en t i o nally blank The YSI Professional Plus Instrument (Pro Plus) is warranted for three (3) years from date of purchase by the end user against defects in materials and workmanship, exclusive of batteries and any damage caused by defective batteries. Pro Plus field cables are warranted for two (2) years from date of purchase by the end user against defects in material and workmanship (6 months for non-field rugged cables*).
* The warranty period for the non-field rugged cables (605107, 605177, 605108, 605178, 605109, 605179) is listed as 6 months. However, the true “working life” of these sensors may be 3-9 months depending on storage and usage in solutions other than clean aqueous samples. ** The warranty for the ammonium, nitrate, and chloride sensors (605104, 605105, 605106) is listed as 6 months.
i Batteries must be installed in the instrument even if powering the unit via the USB connection. This will retain the correct date and time if the PC is turned off. If the USB power is disconnected and there are no batteries in the instrument, the date and time will need to be reset upon subsequent power on. NOTE - On subsequent battery changes you will have approximately 2 minutes to change the batteries before the clock resets.
1010 dual cable Port 1 Options Port 2 Options pH pH Port 1 Options Port 2 Options DO Port Options ORP ORP pH pH pH or pH/ORP* pH or pH/ORP* ammonium ammonium chloride chloride Polarographic 5560 DO Conductivity/ Temperature sensor only nitrate nitrate ORP ORP Galvanic DO pH or pH/ ORP* pH or pH/ ORP* none (port plug) ammonium ammonium chloride chloride nitrate nitrate none (port plug) Quatro Cable (pn: 605790) * If using a 6051010 cable, a sensor must be installed in port
Figure 2. The image on the left shows a clean, dry sensor being aligned with the bulkhead. On the right, the sensor has been pushed into the bulkhead and is being screwed into place. Figure 3. The sensor(s) will install directly in the cable bulkhead. Once installed, the sensor guard will protect the sensor during sampling (DO cap membrane not shown).
Connecting the Cable to an Instrument Keypad 2 To connect a cable, align the keys on the cable connector to the slots on the instrument connector. Push together firmly, then twist the outer ring until it locks into place (figure 5). This connection is water-proof. 3 1 4 13 5 12 6 11 10 7 8 9 Figure 5. Note the keyed connector. The cable and instrument connectors can only be mated once the keys are properly aligned.
Number Key Main Display Description 6 Right Arrow Use to navigate right in alpha/numeric entry screens. Can be pressed simultaneously with Backlight key to increase display contrast. 7 Down Arrow Use to navigate through menus and to navigate down in alpha/numeric entry screens. 8 Power Press to turn the instrument on. Press and hold for 3 seconds to turn off. 9 Help Press to receive hints & tips during operation. 10 Enter Press to confirm selections, including alpha/numeric key selections.
S ys t em M enu Alpha/Numeric Entry Press System menu items. to access any of the following The System menu will allow you to access the setup options of the instrument including; Date/ Time, GLP, Language, Radix Point, Logging, Auto-Shutoff, Backlight, SW (Software) Version, Serial #, and Unit ID. Any item with [brackets] shows the current setting inside the brackets. For instance, in the example at the left, Radix Point is currently set to [Decimal].
GLP The GLP or ‘Good Laboratory Practice’ file saves detailed information about calibrations. It also includes diagnostic information about the sensors. Calibrations are logged into a file, the GLP, for later review as needed. A single GLP file is utilized to store all calibration records and is capable of storing 500 records. Once the GLP file is full, the instrument will begin to overwrite the oldest record with each new calibration record.
An example of a GLP record (Operation performed is single point % DO Calibration) *** Calibrate – DO% *** Date 02/03/09 MM/DD/YY Time 12:14:57PM 12-hour User ID: Tech 1 Probe ID 08D Method Cal Value: Sensor Value: Sensor Type Membrane Type Salinity Mode Temperature Barometer Calibrate Status DO Air Calibrate 100.00% 5.175155uA Polarographic 1.25 PE Yellow 5.175165 Auto 23.9 °C 731.4 mmHg Calibrated assembly from another, typically by serial number.
Language Check the box for Continuous Mode if you want to log samples continuously at a specific time interval. To set the length of time between logged samples, highlight Interval and press enter. Enter the interval as HH:MM:SS. This interval will display at the top of the screen when you select the Start Logging option in run mode. The Pro Plus can be configured to display all text in English, Spanish, French, German, Portuguese, Italian, Norwegian, Simplified Chinese, Traditional Chinese, or Japanese.
Unit ID Unit ID is used to identify instruments in the Data Manager software program that was included with your instrument. It is also used to identify GLP files, Site Lists, Configuration files, and Data files transferred from the instrument to the PC. The default Unit ID is the instrument’s serial number. To modify the Unit ID, highlight Unit ID, press enter and then use the alpha/numeric entry screen. The character limit is 16 characters.
Setup - Dissolved Oxygen Press Sensor enter. , highlight Setup and press enter. Next, highlight DO and press Enabled allows you to enable or disable the Dissolved Oxygen function. Highlight Enabled and press enter to activate(Q) or deactivate(P) dissolved oxygen. Disable dissolved oxygen if you do not have a dissolved oxygen sensor connected to the instrument. If a sensor is Enabled that isn’t connected to the instrument, the display will show an unstable, false reading, ?????, or ----- next to the units.
Local DO allows for localized DO% measurements. This sets the calibration value to 100% regardless of the altitude or barometric pressure. Highlight Local DO and press enter to enable (Q) or disable (P) this function. Local DO is a method for the Pro Plus to factor in the barometric pressure on each DO measurement. In essence, if the barometric pressure changes you wouldn’t notice the difference in the DO% readings in air-saturated water or water-saturated air. Local DO is ideal for EU compliance.
Auto Stable and Hold All Readings enabled, then the display will hold the readings once DO and pH have both reached their Auto Stable settings.
After selecting your User ID and/or Probe ID if appropriate, highlight DO and press enter. Calibrating DO% in Water Saturated Air: 2-Point Calibration with Zero Solution Highlight DO % and press enter to confirm. Place the sensor in a solution of zero DO. The instrument will use the internal barometer during calibration and will display this value in brackets at the top of the display. Highlight Barometer and press enter to adjust it if needed.
Highlight Calibration value and press enter to manually input the sample’s dissolved oxygen value. Highlight Accept Calibration and press enter once the temperature and Dissolved Oxygen readings stabilize. Or, press Esc to cancel the calibration. If User Field 1 or 2 are enabled in the GLP menu, you will be prompted to select the fields after selecting Accept Calibration. After making your selection, press Cal to complete the calibration.
Highlight Calibration Value and press enter to manually enter the correct “true” barometric pressure. Next, highlight Accept Calibration, and press enter. If User Field 1 or 2 are enabled, you will be prompted to select the fields and then press Cal to complete the calibration or press Esc to cancel the calibration. i Laboratory barometer readings are usually “true” (uncorrected) values of air pressure and can be used “as is” for barometer calibration. Weather service readings are usually not “true”, i.e.
Display - Conductivity Press Sensor , highlight Display and press enter. Highlight Conductivity and press enter. Highlight Sp. Conductance (Specific Conductance), Conductivity, Salinity, TDS, or Resistivity, and press enter to select the reporting units for each parameter. One reporting unit per parameter may be enabled. To disable a parameter, select None. You will not be able to display any of these parameters unless the Conductivity sensor is Enabled in the Sensor Setup menu first. Sp.
Choose the units in either SPC-us/cm, C-us/cm or SPC-ms/cm, C-ms/cm and press enter. Calibration - Conductivity i The 6051030 ISE/conductivity cable has a specialized calibration container that resembles a large test tube. This calibration chamber can be used to calibrate the conductivity sensor with an ISE sensor installed. A ring-stand should be used to support this chamber. Highlight Calibration value and press enter to input the value of the calibration standard.
pH Display - pH pH sensors can be used on 60510-X, 6051020-X, 6051030-X, 6051010-X, and Quatro cables. Press Sensor enter. If using a 605103 pH/ORP combination sensor on a 6051020 or 6051030 cable you can report both pH and ORP by configuring ISE1 as pH and ISE2 as ORP in the Sensor Setup menu. Highlight ISE (pH) and press enter. You will not be able to Display the sensor unless it is Enabled in the Sensor Setup menu.
Sensitivity selected by User % Data Variance Threshold 100 - Most Sensitive, Sensitivity bar is set to the far right 0.025% 75 0.39375% 50 1.5% 25 1.13125% 0 - Least Sensitive, Sensitivity bar is set to the far left 0.15% in the Sensor Setup menu may be incorrect. If necessary, highlight the Calibration Value and press enter to input the correct buffer value. Once the pH and temperature readings stabilize, highlight Accept Calibration and press enter to accept the first calibration point.
i The actual readings displayed during the calibration will NOT reflect the updated calibration information. These values will not change until Cal is pressed to finalize the calibration and to update the instrument. i If a sensor is Enabled that isn’t connected to the instrument, the display will show an unstable false reading, ?????, or ----- next to the units. Display - ORP ORP Press Sensor enter. ORP sensors can be used on 60510-X, 6051020-X, 6051030-X, 6051010-X, and Quatro cables.
Sensitivity selected by User % Data Variance Threshold 100 - Most Sensitive, Sensitivity bar is set to the far right 0.05% 75 0.62525% 50 1.275% 25 1.8875% 0 - Least Sensitive, Sensitivity bar is set to the far left 2.5% Within the Auto Stable menu, you can also choose to Hold All Readings for as many parameters as you set for Auto Stable.
Enabled allows you to enable or disable the ISE function and select which ISE sensor is installed on the cable. Highlight Enabled and press enter to enable (Q) or disable (P) the ISE you selected previously (either ISE1 or ISE2). After enabling the ISE function, choose the parameter you want enabled for that ISE. In this example, NH4 is selected.
Calibration - Ammonium, Nitrate, Chloride The 6051030 ISE/conductivity cable has a specialized calibration container that resembles a large test tube. This calibration chamber can be used to calibrate the ISE sensors with the conductivity sensor. A ring-stand should be used to support this chamber. i The ISE sensors can be calibrated at 1, 2, or 3-points. A 2-point calibration without chilling a third calibration solution is extremely accurate and is the preferred method.
After selecting your User ID and/or Probe ID if appropriate, highlight Ammonium, Nitrate, or Chloride to access the appropriate calibration, and press enter. The parameter you want to calibrate may appear under ISE1 or ISE2 depending on your cable type and setup. The message line will show the instrument is ready for the 1st calibration point. The instrument will display the calibration value at the top of the screen. If necessary, highlight the Calibration value and press enter to input the correct value.
the sky and then twist tying, zip tying, or rubber banding the bulkhead to the cable. Making a simple curve to the cable without bending or breaking the cable will allow you to lower the sensor into the aeration tank while the sensor points skyward so the bubbles are no longer bursting on the membrane surface. Conductivity The conductivity sensor will provide quick readings as long as the entire sensor is submerged and no air bubbles are trapped in the sensor area.
time qualifier you would like to set, press enter, and use the numeric entry screen to select the date/time you wish to view. D a t a M anager D esk t o p S o f t ware After making your selections in the Data Filter, highlight Show Data and press enter. The data will have date and time stamps. You will likely have to scroll up and down and side to side using the arrow keys to completely view the data file. No more than 100 data records can be viewed at one time.
Using the Communications Saddle u WARNING: DO NOT connect the Communications Saddle to your PC before installing Data Manager. The Communication Saddle drivers must be installed prior to connecting it to your PC. The drivers will install automatically during the Data Manager installation. The first time the saddle is connected to the PC, you may have to walk through a couple of installation wizards.
Configure Instruments Data Manager allows for easy and quick configuration of single or multiple instruments. Once you have uploaded a site list or configuration file, you can edit it as needed, save it, and download it to other instruments. You no longer need to configure each instrument individually. By using the same configuration file for all instruments, you can rest assured that all instruments will have identical settings. New site lists and configuration files can be created in Data Manager as well.
General Maintenance - Sensor Ports Thread the membrane cap onto the sensor. It is normal for a small amount of electrolyte to overflow. It is important that the entire sensor connector end be dry when installing, removing or replacing. This will prevent water from entering the port. Once a sensor is removed, examine the connector inside the port. If any moisture is present, use compressed air to completely dry the connector or place directly in front of a steady flow of fresh air.
cap. If the cathode remains tarnished, contact YSI Technical Support or the Authorized dealer where you purchased the instrument. Silver Anode After extended use, a thick layer of Silver Chloride (AgCl) builds up on the silver anode reducing the sensitivity of the sensor. The anode must be cleaned to remove this layer and restore proper performance. The cleaning can be chemical or mechanical: Chemical cleaning: Remove the membrane cap and rinse the electrodes with deionized or distilled water.
Sensor Maintenance - Conductivity The openings that allow sample access to the conductivity electrodes should be cleaned regularly. The small cleaning brush included in the Maintenance Kit is ideal for this purpose. Dip the brush in clean water and insert it into each hole 10 to 12 times. In the event that deposits have formed on the electrodes, it may be necessary to use a mild detergent (laboratory grade soap or bathroom foaming tile cleaner) with the brush.
sensor can be regenerated by washing with alcohol and/or gently polishing with fine emery paper in a circular motion to remove any deposits or discoloration, then thoroughly washing with deionized water to remove any debris. The sensor may require soaking in the high standard chloride calibration solution to recover its performance. Long-term Storage - Conductivity No special storage is required.
Long-term Storage - ORP Error Messages To store, remove the sensor from the cable and seal the vacant port with the provided plug. Fill the original shipping/storage vessel (plastic boot or bottle) with buffer 4 solution and then submerge the sensor into the solution. The sensor should remain submerged in the solution during the storage period; therefore, make certain that the vessel is sealed to prevent evaporation and periodically check the vessel to ensure the sensor does not dry out.
A ccess o ries / P ar t N umbers Dissolved Oxygen The dissolved oxygen sensors will use Probe Current (DO uA) and Probe Slope (%/uA) as part of their GLP file records.
Sensor Part Number Description Accessory Part Number Description 605105**** Chloride (ISE) 605990 Flow cell kit for Quatro cable assemblies.
Accessory Part Number 603069 063517 063507 Description Belt clip Ultra clamp Solutions Part Number Description 3161 1,000 us/cm conductivity solution (quart) 3163 10,000 us/cm conductivity solution (quart) 3169 50,000 us/cm conductivity solution (8 pints) 3682 Zobell ORP solution (125 mL) 3824 pH 4, 7, 10 buffers (2 pints of each) 3841 1 mg/L ammonium solution (500 mL) 3842 10 mg/L ammonium solution (500 mL) 3843 100 mg/L ammonium solution (500 mL) 3885 1 mg/L nitrate solution (500 m
Harmonized Standards: Supplementary Information: Authorized EU Representative • EN61326-1:2006, Electrical equipment for measurement, control, and laboratory use – EMC requirements – Part 1: General Requirements • EN61326-2-3:2006, Electrical equipment for measurement, control and laboratory use – EMC requirements – Part 2-3: Particular Requirements – Test configuration, operational conditions, and performance criteria for transducers with integrated or remote signal conditioning.
The undersigned hereby declares on behalf of the named manufacturer under our sole responsibility that the listed product conforms with the Australian and New Zealand Electromagnetic Compatibility (EMC) requirements for generic products to be used in residential, commercial, and light industrial environments. Manufacturer: Product Name: • Batteries are removed and sent to battery recyclers for dedicated metals. When the time comes for you to recycle, follow the easy steps outlined at www.ysi.com.
Item # 605596 Rev D Drawing # A605596 March 2009 ©2009 YSI Incorporated.