Pro1030 U S ER MA NUAL E nglish
C o ntents Warranty................................................................................................. i Introduction..........................................................................................1 Getting Started.....................................................................................1 Initial Inspection......................................................................1 Battery Installation..................................................................1 Key Pad..
Factory Default......................................................................14 Calibration ..........................................................................................15 Temperature..........................................................................15 pH Calibration.......................................................................15 ORP Calibration.....................................................................17 Conductivity Calibration................................
I ntr o ducti o n T his page left intenti o nally blank Thank you for purchasing the YSI Pro1030, an instrument from the YSI Professional Series product family. The Pro1030 measures conductivity, temperature and either pH or ORP in water. The Pro1030 features an impact resistant and waterproof (IP-67) case, a rugged MS-8 (militaryspec) cable connector, backlit display, user-selectable sensor options, 50 data set memory and a rubber over-mold case.
4. Install the new batteries, ensuring correct polarity alignment (figure 1). 5. Place the battery cover on the back of the instrument and tighten the four screws. Do not over-tighten. Number i Key Pad 2 3 1 4 Description 1 Calibrate Press and hold for 3 seconds to calibrate. Opens Calibrate menu from the Run screen. 2 Up Arrow Use to navigate through menus, to navigate through box options along the bottom of the Run screen and to increase numerical inputs.
i When an ISE sensor is not installed in the cable, the bulkhead connector is not water-proof. Do not submerge the cable without a sensor installed. Submerging the cable without a sensor installed may cause permanent damage to the cable that is not covered under warranty. Connecting Instrument the Probe/Cable Assembly to the To connect the cable, align the keys on the cable connector to the slots on the instrument connector.
Backlight First Power On Once the instrument is powered on, pressing the power/backlight key The instrument will step through an initial configuration when powered on for the first time. This will set the language. Use the up or down arrow keys to highlight the appropriate language, then press enter to confirm (figure 6). will turn on the display backlight. The backlight will remain on until the key is pressed again or after two minutes of not pressing any key on the keypad.
screen. Highlight the desired unit (Celsius or Fahrenheit) and press enter to enable. The enabled temperature unit will have an ‘X’ in the box next to it. Only one unit may be enabled at a time. Highlight the ESC-Exit box and press enter to save any changes and to close the Temperature Units submenu. ISE Sensor Type ISE Sensor Type sets the type of ISE sensor being used; either pH (model #1001) or ORP (model #1002). Figure 7, page 1 of System Setup menu.
When Auto Stable is enabled, an AS symbol will display next to the reading on the Run screen and blink during stabilization. When the ISE and/or conductivity reading stabilizes based on the Auto Stable settings, the AS symbol will display steadily and the instrument will beep twice if Audio is turned on. ISE Auto Stable can be set to a % change of 0.0 to 9.9% over 3 to 19 seconds.
Using the Pro1030’s default reference temperature and temperature coefficient (25 °C and 1.91%), the calculation is carried out as follows: Specific Conductance (25°C) = Conductivity of sample 1 + 0.0191 * (T - 25) T = Temperature of the sample in °C Specific Conductance Reference Temperature (SPC Ref. Temp.) SPC Ref. Temp. (Specific Conductance Reference Temperature) is the reference temperature used to calculate Specific Conductance. The reference temperature range is 15 and 25°C.
Next, use the up or down arrow keys to adjust the shut off time from 0 to 60 minutes. Press enter to save the new shutoff time. Next, highlight the ESC-Exit box and press enter to close the submenu. Conductivity Auto Stable To disable Auto Shutoff, set the Time in Minutes to 0 (zero). SPC Reference Temperature 25°C SPC Temperature Coefficient 1.91%/°C Cell Constant Parameter Reset Defaults Off (0.
7. ‘Calibration Successful’ will display for a few seconds to indicate a successful calibration and then the instrument will return to the Run screen. 8. If the calibration is unsuccessful, an error message will display on the screen. Press the Cal key to exit the calibration error message and return to the Run screen. See the Troubleshooting guide for possible solutions. 2-Point Calibration 1. Place the sensor in pH buffer 7 or 6.86 and allow the temperature and pH readings to stabilize. 2.
5. Wait for the temperature reading to stabilize, then press enter to complete the calibration or press Cal to cancel. 6. ‘Calibration Successful’ will display for a few seconds to indicate a successful calibration and then the instrument will return to the Run screen. 7. If the calibration is unsuccessful, an error message will display on the screen. Press the Cal key to exit the calibration error message and return to the Run screen. See the Troubleshooting section of this manual for possible solutions.
7. If the calibration is unsuccessful, an error message will display on the screen. Press the Cal key to exit the calibration error message and return to the Run screen. See the Troubleshooting section of this manual for possible solutions. Calibrating in Salinity 1. Place the sensor into the solution. The solution must cover the holes of the conductivity sensor that are closest to the cable (figure 8).
Viewing Data Once in Data mode, use the up and down arrow keys to view saved data sets in sequential order or press enter to access the bottom functions. After accessing the bottom functions, highlight the Data box and press enter to regain access to viewing data. The data set displayed is indicated by the data set number, 01 through 50. Erasing Data While viewing saved data, press the enter key to access the function boxes at the bottom of the display.
13, and 15 becomes 14. The instrument will return to Data mode after erasing one data set. Select Erase All Data Sets and press enter to clear the Pro1030 memory and return to Data mode. from entering the port. Once the ISE sensor is removed, examine the connector inside the port. If any moisture is present, use compressed air to completely dry the connector or let it air dry. If the connector is corroded, contact YSI Technical Support or the YSI authorized dealer where you purchased the instrument.
1. Soak the sensor for 10-15 minutes in clean water containing a few drops of commercial dish washing liquid. 2. GENTLY clean the glass bulb and platinum button by rubbing with a cotton swab soaked in the cleaning solution. 3. Rinse the sensor in clean water, wipe with a cotton swab saturated with clean water, and then rerinse with clean water.
Symptom Instrument will not calibrate pH or ORP; instrument displays “Calibration Over”, “Calibration Under”, or “Unstable Reading” during calibration. Possible Solution 1. 2. 3. 4. 5. pH or ORP readings are 1. inaccurate. 2. 3. 4. 5. pH values display Over or Undr on Run screen. 1. 2. 3. 4. 5. 6. ORP values display Over or Undr on Run screen. 1. 2. 3. 4. 5. 6. Verify correct sensor type selection in the System Setup menu. Verify the calibration solution is accurate.
S pecificati o ns A ccess o ries / P art N umbers These specifications represent typical performance and are subject to change without notice. For the latest product specification information, please visit YSI’s website at ysi.com or contact YSI Tech Support. Parameter Range Resolution Accuracy Part Number Description 6051030 Pro1030 Instrument 6261030-1, -4, -10, -20, or -30 1, 4, 10, 20, 30-meter cable assembly* (3.2, 13, 32.8, 65.6, 98.4-feet) -5 to 55°C 0.1°C ± 0.
D eclarati o n o f C o nf o rmity The undersigned hereby declares on behalf of the named manufacturer under our sole responsibility that the listed product conforms to the requirements for the listed European Council Directive(s) and carries the CE mark accordingly.
When placing an order please have the following available: 1.) YSI account number (if available) 2.) Name and phone number 3.) Purchase Order or Credit Card number 4.) Model Number or brief description 5.) Billing and shipping addresses 6.) Quantity Service Information YSI has authorized service centers throughout the United States and Internationally. For the nearest service center information, please visit ysi.