User manual
Relay outputs System 182-XT-4
6 - 2
ba76027e02 12/2014
6.2 Linking and adjusting: general procedures
6.2.1 linking outputs
Linking options You have the following options of linking relay outputs:
Sensor Sxx
The output is linked with sensor 0x
Sensor S01&S02
The output is linked to both sensors S01 and S02.
System
Select this option if you want to use the output for system monitoring
(details, see section 6.4.2).
Procedure:
linking outputs
Proceed as follows if the output was not yet linked. If you want to
change a link, you have to erase the existing link first.
1 Open the SETTINGS menu with s.
2 Using
w and g, select and confirm the Relay output Rx
menu item (corresponding to sensor 1, 2 or 3). A list with all
linking options is displayed.
3 Select the required option with w and confirm with g.
The output is linked. The link is displayed in the header when
the setting of the output is edited.