User manual

Operation System 182-XT-4
4 - 16
ba76027e02 12/2014
Putting the sensor into
operation after
calibrating
5 Confirm the calibration data with g. Calibration is completed
with this. The following display message describes the further
steps to put the sensor into operation again.
6 Confirm with g. The display returns to the measured value
display (the measured value flashes as the sensor is still in the
maintenance condition).
7 If calibration was successful, immerse the sensor in the test
sample.
8 Wait for a stable measured value.
9 Switch off the maintenance condition (highlight sensor and
press
g three times). The measured value has stopped
flashing.
If calibration was not successful, this is indicated by "----"
on the display and a corresponding message with remedial
actions appears in the log book. Follow the instructions and
repeat calibration.