User manual
System 182-XT-4 Installation
3 - 33
ba76027e02 12/2014
– marked as an interrupt facility for the DIQ/S 182-XT-4-PR.
After it has been installed, the DIQ/S 182-XT-4-PR may only be
opened if the mains voltage has been switched off beforehand.
Materials required Wire end sleeves, suitable for the power line, with suitable crimping
tool
1 x cable gland, matched to the cable diameter (see section 3.7 on
page 3-31).
Tools Cable stripping knife
Wire stripper
Phillips screwdriver
Small screwdriver.
Preparing the power
cable
Fig. 3-25 Prepared power cable.
NOTE
The ground wire must not project into the enclosure. Otherwise, mal
-
functions could occur.
1 Cut off the cable to the required length.
2 Strip the cable insulation for approx. 45 mm.
3 Bare the wires of phases L and N and fit them with wire end
sleeves.
4 If present, cut off the ground wire at the end of the cable
sheath.