User manual

System 182-XT-4 Installation
3 - 31
ba76027e02 12/2014
3.7 Electrical connections: General instructions
Cable glands All electric cables are fed from below via prepared openings in the
enclosure of the DIQ/S 182-XT-4-PR and the MIQ modules. Cable
glands with different clamping ranges are included with the DIQ/S 182-
XT-4-PR to provide sealing between the cable and enclosure as well as
for strain relief. Select the matching cable gland for the respective cable
diameter:
Small, clamping range 4.5 to 10 mm. This cable gland is suitable for
all IQ S
ENSORNET cables (including earth cable after stripping the
outer insulation, see section 3.4.4) and IQ S
ENSORNET sensor
connection cable.
Large, clamping range 7 to 13 mm. This cable gland is required for
cable sheaths with an outside diameter of more than 10 mm and is
screwed into the enclosure via an extension piece.
If necessary, you can order other sizes of cable gland (see
chapter 11 A
CCESSORIES AND OPTIONS).
sealing ring 20 x 15 x 1 mm
cable gland M16
blind plug
sealing ring 20 x 15 x 1 mm
extension piece M16/M20
sealing ring 24 x 19 x 2 mm
cable gland M20