Instruction Manual

7
ENGLISH
Measurement
Fixtheadapterfor16mmvialsonthesamplechamber.
Turntheunitonusingthe[ON/OFF]key.
Thedisplayshowsthefollowing:
Selecttherequiredtestusingthe[MODE]key.
Scroll Memory (SM)
Toavoidunnecessaryscrollingfortherequiredtestrange,theinstrument
memorizes the last rangeusedbeforebeingturnedoff.Whentheinstrumentis
turned on again, the scroll list comes up with the last used test range.
Thedisplayshowsthefollowing:
Placetheblankintheadapter(Note14)makingsurethatthe
marks
are aligned.
Blanksarespeciallypreparedforeachindividualtestrange.
Pressthe[ZERO/TEST]key.
The“Range“symbolashesforapprox.8seconds.
Thedisplayshowsthefollowing:
Afterzeroingremovethevialfromtheadapter.
Placethesampleintheadapter(Note24)makingsurethatthe
marks
are aligned.
Pressthe[ZERO/TEST]key.
The“Range“symbolashesforapprox.3seconds.
The result appears in the display.
LR-/MR-range: in mg/l
HR-range: in g/l
The result is saved automatically.
Tolerance:±3.5 %(fullscale)
Repeating the test:
Pressthe[ZERO/TEST]keyagain.
New zero calibration:
Pressthe[ZERO/TEST]keyfor2seconds.