Instruction Manual

6
Methods
COD Reagents
Selecttheappropriatevialforthedesiredrange*:
LR-Range: 0–150mg/l,partnumber251990Y,25count
MR-Range: 0–1500mg/l,partnumber251991Y,25count
HR-Range: 0–15000mg/l,partnumber251992Y,25count
*Reagents are also availble in 150 count packages.
Preparing the Sample
Openareactionvialwithawhitecapandaddthespeciedvolume.
(Ensurethatappropriatepersonnelsafetyequipmentisused.)
LR-/MR: 2 ml water sample
HR: 0.2 ml water sample
Prepareablank(Note1)byusingdeionizedwater(TOC-free)insteadofthesample
(LR/MR: 2 ml, HR: 0.2 ml).
Replacethecaptightly.Invertthevialgentlyseveraltimestomixthecontents(Thevialwillbecome
hotduringmixing!)anddigestthevialsfor120minutesinthereactoratatemperatureof150°C.
Removethevialsfromthereactorandallowthemtocooldownto60°Corless.Mixthecontentsby
invertingeachvialseveraltimeswhilestillwarm.Thenallowthevialstocooltoambienttemperature
beforemeasuring.
Guidelines for photometric measurements
1. Runsamplesandblankswiththesamebatchofvials.
Theblankisstablewhenstoredinthedarkandcanbeusedforfurthermeasurementswithvials
fromthesamebatch.
2. Don‘tplacehotvialsintheadapter.Allowthevialstocooltoroomtemperaturefor
a minimum 45 minutes. It is recommended to leave the vials to cool over night.
3. Suspended solids in the vial lead to incorrect measurements. For this reason it is important to place
thevialscarefullyintheadapter.Theprecipitantatthebottomofthesampleshouldnotbe
suspended.
4. Cleantheoutsideofthevialswithatoweltoremovengerprintsorothermarks
.
5. Avoidspillageofwaterorreagentsolutionintothesamplechamberbecausethiscanleadtoincor-
rect test results.
6. Contaminationofthetransparentcellchambercanresultinincorrectreadings.Checkatregularin-
tervalsand–ifnecessary–cleanthetransparentcellchamberusingamoistclothorcottonswab.
7. Largetemperaturedifferencesbetweentheinstrumentandtheenvironmentcanleadtoerrors–
e.g.duetotheformationofcondensationinthecellchamberoronthevial.
8. Toavoiderrorscausedbystraylightdonotusetheinstrumentinbrightsunlight.