Instruction Manual

12
Technical Data
Instrument: double wavelength, automatic wavelength selection,
direct reading colorimeter
Light source: LEDs,interferencelters(IF)andphotosensorintransparent
cellchamber.WavelengthspecicationsoftheIF:
430 nm
= 5 nm
610 nm = 6 nm
Wavelength accuracy: ± 1 nm
Photometric accuracy*: 3%FS(T=20° C–25° C)
Photometric resolution: 0.01 A
Power supply: 4batteries(AAA/LR 03)
lifetime:approx.17hoursor5000tests
Auto-OFF: instrumentautomaticallyturnsoff 10minutesafterlastkeypress
Display: backlitLCD(onkeypress)
Storage: internalringmemoryfor16datasets
Serial Interface: IRinterfacefordatatransfer
Time: realtimeclockunddate
Calibration: userandfactorycalibration
resettingtofactorycalibrationpossible
Dimensions: 155 x 75 x 35 mm (LxWxH)
Weight: approx.260g(incl.batteries)
Ambient conditions: temperature:540°C
rel. humidity: 3090% (non-condensing)
Waterproof: oating;IP68(1hourat0.1meter)
CE: CerticateforDeclarationofCE-Conformity
*measuredwithstandardsolutions
Toensuremaximumaccuracyoftestresults,alwaysusethereagentsystems
suppliedbytheinstrumentmanufacturer.