Reset Systems to factory defaults, and test Relay(s).
Reset Clean Probes
Menus → System → Service→ Reset Clean Probes
Enable clean probe timer, congure the duration, and reset clean probe days remain-
ing from Reset Clean Probe menu. Selecting reset at this menu resets clean probe
timer to the user congured value.
icon displays at Run Screen when the
clean probe timer expires.
Reset Clean Probe timer at the
menu or at the Reset Clean Probes menu.
When prompted “Are you sure?”, select “Yes” to reset the clean probe timer to the
Conguring the 5X00
user congured number of days.
Reset To Factory
Menus → System → Service→ Reset To Factory
e Reset to Factory menu allows resetting of some or all of 5X00 systems to factory
defaults. Example - if Clean Probe Timer is user congured to 60 days, performing a
factory reset will reset the Clean Probe Timer to 90 days which is the factory default
setting and not the user congured value. See Appendix B Menu Maps starting on
page 205 for factory default settings for all systems.
Notes:
- “Reset All” and “Reset All Systems” resets time format to 12 hour and date
format to MMDDYYYY.
- It is strongly recommended that user conguration be maintained in a log
by the user.
- Instrument, aux, sensor, and sux names are not reset at any reset to factory
menus. Relay names are reset at “reset relays” and “reset all” menus.
- Control icons and control menu messages are displayed for 30 seconds aer
pressing
from reset to factory menus for any system that was active
(energized relay) prior to factory reset.
Reset All
Menus → System → Service→ Reset To Factory→ Reset All
Resets all systems, sensors, and timers to factory defaults.
1.
Use
▲
and
▼
to scroll and highlight submenus.
2.
Press
to select.
3. When prompted, “Are you sure?”, select one of the following to conrm,
Conguring the 5X00