Manual

YSI 5X00
117
YSI 5X00
116
1. Enter Display Settings menu
Menus → System → Display Settings →Display Format
2.
Use
and
to scroll and highlight submenus.
3.
Press
to select.
4.
to save conguration and return to Run Screen .
Notes:
- At 1x2 & 2x2 display settings, analog aux values over 7 digits will overlap on to
the adjoining sensor window. Adjust aux display format to lower resolution.
- Because of font limitations, 1X2 and 2X3 display formats do not show all
20 characters of a 20 character name. Typically only 13 characters are dis-
played.
ODO Cal K (5500D-01, 5500D-02 and 5500D-04)
Menus → System → ODO Cal K →DO1-DO4
e ODO Cal K system menu appears only in 5500D-01, 5500D-02 and 5500D-04
instruments. is menu is for viewing and uploading DO sensor cap calibration
codes. Uploading ODO sensor cap calibration codes is required whenever a sensor
cap is replaced.
Cap calibration codes are not stored in the 5500D instrument but rather are stored
5500
Conguring the 5X00
in the ODO sensor. Upon selecting DO-1, DO-2, DO3 or DO-4 at the ODO Cal
K menu communication initiates between the instrument and the sensor. An ODO
sensor has six cap calibration codes. ey are known as K1, K2, K3, K4, K5 and KC.
Direction to view or upload ODO sensor cap calibration codes:
1. Enter ODO Cal K menu
Menus → System → ODO Cal K →DO-1, DO-2, DO-3, DO-4
2. Select DO -1, DO-2, DO-3 or DO-4
ODO sensor cap calibration codes are displayed
If the 5500D is unable to obtain the cap calibration codes from the ODO sensor the
below error message is displayed. Typically a retry aer several seconds will display
the ODO sensor cap calibration codes. If the calibration codes cannot be displayed
conrm that the ODO sensor is displaying a value at the run screen.
3. To upload sensor cap calibration codes select K1, K2, K3, K4, K5 or KC.
4. Use alpha/numeric keyboard to enter the values exactly as provided on
the sensor cap instruction sheet sent from YSI.
Conguring the 5X00