User Manual

13
Symptoms Actions
Does not hold
calibration
Allow sufficient time for temperature and
oxygen stabilization before calibration
Check for water droplets on membrane
Check membrane for a pinhole or split
Change membrane
Clean probe (see instructions under
Electrode Cleaning)
Unstable readings Check membrane for a pinhole or split
Allow 60 sec. for reading to stabilize
Change membrane
Clean probe (see instructions under
Electrode Cleaning)
Membrane cap is
tight
Use pliers to loosen the cap (turn counter
clockwise while facing sensor tip)
S
ERVICING
The 5239 probe has only two replaceable parts (other than the
membrane cap), the probe guard and the 5-pin MS connector
on the end of the cable. The cable is permanently attached to
the probe and cannot be replaced.
YSI part number Description
110275 5-pin MS connector kit
059956 Probe guard
5238 Probe reconditioning kit