Troubleshooting
Network Troubleshooting
Issue possible cause possible solution
no communication to
networked unit(s)
Are master/slave menus
properly congured?
See Networking - page 93.
Is termination relay en-
abled in 1st & last 5200A on
network?
See Termination Relay -
page 95.
Is proper communication
method congured and
wired properly?
See Networking - page 93.
See Wire Communication
Equipment - page 57.
Are slaves enabled and recog-
nized at master?
enable slaves at Menus →
System → Communications
→ Network →Master
more than one device is
congured as a master
verify conguration at
Menus → System → Com-
munications → Network
→Master/Slave menus on
all nodes.
slow network perfor-
mance
nicked wires;
improper wiring
Verify network wiring -
See Wire Network - page 52
5200A Systems Troubleshooting
Issue possible cause possible solution
time and date lost at
power up
bad lithium battery replace lithium battery
no events in event log event logging not enabled See Event Logging - page
104.
no data in data log data log interval time has not
passed
See Data Logging - page102.
cannot view specic data
log records at 5200A
Data log record has older
date and time-stamp than
the 25 most current data log
records available at 5200A
sensor log.
See Data Logging - page102.
Troubleshooting
Firmware Updater Troubleshooting
Issue possible cause possible solution
Device “timed out” error incorrect conguration in
5200A communication menu
See Device Timed Out er-
ror - page 112.
incorrect wiring of RS232
connection
See Device Timed Out er-
ror - page 112.
5200A does not go
into downloader mode
when “program device
is selected at Firmware
Updater
Manually put 5200A in
downloader and retry rm-
ware updater upload.
Menus → System → Down-
loader
See Force Downloader -
page 113.
unit does not upload new
code at downloader
unit is not in downloader
mode
See Downloader - page 109..
RS232 connection is incor-
rect
See Wire Communications -
RS232 - page 58.
Feed Timer Troubleshooting
Issue Understanding how the 5200A calculates sensor reduction
time
Sensor reduction % is
not accurate based on
displayed sensor value(s).
e 5200A displays temperature sensor values to the 10ths
and DO values to 100ths. In reality the 5200A temperature
and DO sensor system values are stored at a much greater
resolution (out to 64 bits). It is the stored values, not the dis-
played values, that are used to calculate the feed timer sensor
reduction.