Press 6. to save conguration and return to Run Screen.
Notes:
To congure a negative set point, control, and alarm values enter “-” aer, -
not before, the numeric digit(s).
See 30 Second Sensor Hold O - - page132.
Only whole numbers can congured for ORP set point, control, and -
alarm values.
pH
Menus →Sensor Setup→pH
See Set Points, Control, and Alarms - page 128 before conguring the pH system.
Timed and Normal Control
pH high and low control can operate in two dierent modes, normal or timed. In the
normal mode, control relays are energized until the set point value is reached. e
normal mode can be used for either low control, high control, or both. For normal
control, congure dose up/dose down to 0. e dwell time will be ignored.
In the timed mode, an energized control relay is energized for the dose time, re-
gardless if the sensor value has returned to the set point value. Aer the dose time
expires, the control relay de-energizes for the congured dwell time. Aer the dwell
time expires and if the value is still in control range, the control relay will reenergize.
A dwell time must be congured in the timed mode or the control relay will never
de-energize regardless of sensor value.
Conguring the 5200A
Directions to Congure pH sensor
Connect probe assembly - 1. page 60.
Enter pH Sensor Setup menu2.
Menus →Sensor Setup→pH
Use3.
▲
and
▼
to scroll and highlight submenus.
Press4.
to select.
Congure submenus.5.
Press 6. to save conguration and return to Run Screen.
Notes:
pH set point, control, and alarm values can be up to ve (5) characters, -
including the decimal point. Values truncate to 100ths when no 10s
digit is congured.
See 30 Second Sensor Hold O - - page132.
Active dosing timers are reset aer systemwide events. Systemwide events -
de-energize relays. Example - pH dose up relay is congured for two (2)
minutes. Energized dose up relay de-energizes when 5200A goes into Svr
Hold. e dose up relay had been energized for 15 seconds before the
Svr Hold occurs. Svr Hold time is one (1) minute. Aer Svr Hold time
expires, the control system is reset. If the sensor is reporting a control
low condition, the dose on time will be for two (2) minutes and not the
remaining 45 seconds of the dose up time prior to Svr Hold.
pH Control Low Range message remains at the
- menu when
the pH system is disabled and when a control low relay is energized.
Power cycle the 5200A to clear the control low range message at the
Msg’s menu.
Conguring the 5200A