5561 probe/cable users should ALWAYS disable the conductivity system -
since there is no conductivity sensor in use to accurately compute a
salinity value.
Changes to salinity aect the DO mg/L value. -
Changes to salinity do not aect the DO %Sat value. -
e DO system uses the temperature sensor value regardless if tempera- -
ture is enabled. It is important to verify that the temperature sensor is
measuring accurately.
Two Range Controls for DO
e 5200A DO system has two low and two high control ranges. is dual range
allows for more control over oxygen levels in the system stream. Example - DO
range low one system becomes active, an air injection system energizes when properly
wired and congured to relay #1. e DO level continues to fall. Range low two
system becomes active and a pure oxygen injection system energizes when properly
wired and congured to relay #2.
Directions to Congure DO Sensor
Connect probe assembly - 1. page 60.
Enter DO Sensor Setup menu2.
Menus →Sensor Setup→DO
Use3.
▲
and
▼
to scroll and highlight submenus.
Conguring the 5200A
Press
4.
to select.
Congure submenus.5.
Press 6. to save conguration and return to Run Screen.
Notes:
DO set point, control, and alarm values are always entered in mg/L re- -
gardless of display format.
DO set point, control, and alarm values can be up to three (3) characters, -
excluding decimal point. Values will truncate to 10ths when 10s digit
is congured.
5200A calculates the correct %Sat value based on configured elevation -
during calibration.
When DO range low 2 or DO range high 2 are active with energized -
relays, only the range low 2 and range high 2 control messages and icons
display at the control menu and Run Screen regardless if DO range low
1 or DO range high 1 relays are also energized.
See 30 Second Sensor Hold O - - page132.
ORP
Menus →Sensor Setup→ORP
See Set Points, Control, and Alarms - page 128 before conguring ORP system.
Directions to Congure ORP sensor
Connect probe assembly - 1. page 60.
Enter ORP Sensor Setup menu2.
Menus →Sensor Setup→ORP
Use3.
▲
and
▼
to scroll and highlight submenus.
Press4.
to select.
Congure submenus.5.