User Manual

YSI 5200A
12
YSI 5200A
13
Figure 1.1
Preview of User Manual
is manual includes information on installing, conguring, operating, and main-
taining a 5200A system.
Read Chapter 2 “Safety Information before installing or conguring a 5200A
system.
Chapter 3 “Installation provides information on how to install the 5200A, 5200A
components, and peripheral devices. Only qualied people should install a 5200A
system. Do not attempt to install or congure this device if you are not qualied
to do so. If information is not clear, DO NOT PROCEED until clarication can be
obtained. All wiring involving connections to main power must be performed by
a qualied licensed electrician and must conform to all locally applicable electrical
codes.
Chapter 4 “Conguring 5200A Systemsprovides information on menu structure
and conguring 5200A instruments.
Chapter 5AquaManager” provides general information regarding the PC soware
program. e AquaManager program includes an extensive help system, which
contains detailed information that is not provided in this manual.
Introduction
is icon is used throughout the manual to indicate helpful information for Aqua-
Manager users.
Chapter 6 “Maintenance provides information on maintaining the 5200A system
including the monitor, sensor, and component service.
“Notes at the end of each manual section provide additional information to help
clarify certain 5200A operations and features.
e appendices include information on Unpacking & Inspection, Troubleshooting,
Menu Maps, Declaration of Conformity, Warranty information, and SMS Messag-
ing.
An index is provided starting on page 228.
Product Specications
To ensure accurate and reliable performance, operate the 5200A system according to
specications provided in the following tables. Damage and/or incorrect operation
could result if these specications are not followed.
Product specications are subject to change without notice. Visit www.ysi.com to
view the most recent product specications.
Environmental Specications
Environmental Category Operating Range
Temperature -15 to 70C (5 to 158F)
Enclosure Rating NEMA 4X (direct exposure to outside conditions),
IP65
Compression Fitting Rating IP 68/NEMA 6 (liquid tight)
Electrical Safety IEC compliant*
Other RoHS compliant
* See Appendix 4 Declaration of Conformity - page 218.
Introduction