User Manual
YSI 5200A
136
YSI 5200A
137
Aux Relay Analog
When enabled, the Aux analog relay energizes when either the control/alarm high
OR the control/alarm low value is processed by the 5400. Aux relay energizes aer
the congured aux hold o time. When alarms are disabled and as long as value
≤ control/alarm low value OR value ≥ control/alarm high value is processed by
the 5400, the relay remains energized including when menus are entered. Relay
de-energizes when a value > control/alarm low value OR value < control/alarm
high value is processed by the 5400 regardless if alarms are enabled. If alarms
are enabled, the relay also de-energizes when one of the four front panel sokeys
is pressed. Pressing one of the sokeys resets the
alarm system. e Alarm system becomes active again if, aer the aux hold o
period, the alarm condition still exists.
Aux analog events (event log) are identied by aux number, relay state (on/o), and
report “alarm high” or “alarm low” condition regardless if alarms are enabled.
AquaManager identies Aux analog events as control or alarm based on node
conguration.
Directions to Congure Aux (Analog)
Wire Aux analog devices - 1. page 49.
Select “0-1v”, “0-5v”, or “4/20ma” input type at
2. Menus → System → Con-
gure Aux →Input Type
Enter Aux 1-2 in Sensor Setup menu3.
Menus →Sensor Setup→Aux 1-2
Use4.
▲
and
▼
to scroll and highlight submenus.
Press5.
to select.
Congure submenus.6.
Press 7. to save conguration and return to Run Screen.
Conguring the 5200A
Notes:
Control/alarm high value must be ≤ Max value. Control/alarm low -
value must be ≥ Min value. Min & Max values are congured in Menus
→ System → Congure Aux →Aux 1-2→Min/Max menu.
Do not enable aux analog alarms when the sensor is in the control/alarm -
condition. Enabling aux analog alarms when sensor is in control/alarm
condition can result in aux system not properly processing existing alarm
condition.
Conductivity
Menus →Sensor Setup→Conductivity
e conductivity system operates in one of four ranges. e ranges are: 0-200 µS,
0-2000 µS, 0-20mS, and 0-200mS. e conductivity range is automatically deter-
mined based on the readings from the sensor. Conductivity values >10,000 µS are
displayed in mS. See Set Points, Control, and Alarms - page 128 before conguring
conductivity system.
Notes:
When conductivity is enabled, salinity is ALWAYS calculated by the -
conductivity sensor value. is value is used to calculate the DO mg/L
value. e salinity value is displayed at the Run Screen when salinity is
enabled. See Salinity - page 145.
Disabling conductivity requires that a ppt value be user dened in the -
5200A DO sensor menu.
5561 probe/cable users should ALWAYS disable the conductivity system -
since there is no conductivity sensor to accurately compute a salinity
value.
Conductivity set points do not self adjust based on the range of the sensor. -
See Menu Maps - 194 for default conguration.
Timed and Normal Control
Conductivity high and low control can operate in two dierent modes, normal or
timed. In the normal mode, control relays are energized until the set point value
is reached. e normal mode can be used for either low control, high control, or
both. For normal control, congure the dose up/dose down to 0. e dwell time
will be ignored.
In the timed mode, an energized control relay is energized for the congured dose
time, regardless if the sensor value has returned to the set point value. Aer the
dose time expires, the control relay de-energizes for the congured dwell time.
Conguring the 5200A