User Manual

YSI 5200A
126
YSI 5200A
127
Notes:
- Always manually poll the 5200A aer uploading conguration changes
and verify the uploaded information is retained. When conguration
changes are uploaded using AquaManager at the same time congura-
tion changes are being made by a user in the menus on the 5200A, the
AquaManager uploads may not be saved depending on the timing of
keystrokes at the 5200A.
e Run Screen must be displayed during normal operation. - If the
Conguring the 5200A
5200A is not at the run screen during AquaManager uploading, pressing
at the 5200A will only save certain conguration changes.
See Chapter 5 AquaManager - page 166 and AquaManager online help for additional
information.
Sensor Setup
Menus → Sensor Setup→Aux 1
→Aux 2
→Conductivity
→DO
→ORP
→pH
→Temperature
→Salinity
Conductivity, DO, ORP, pH, and temperature system set points, control, and alarm
relays are congured in the Sensor Setup menu. In addition, the salinity system
is enabled in this menu. Congure control and alarm values in order to maintain
desirable water quality for your water system.
Aux systems operate dierently than 5200A sensor systems. See information begin-
ning on page 133 regarding aux systems setup.
Values and conditions at sensor and aux device locations are processed by the 5200A.
Systems (alarm and/or control) become active and inactive based on user congura-
tion. Four onboard relays can be assigned for control and alarm devices. Enabled
relays energize when a control or alarm system is active. Example - DO sensor (input)
reports a 5.4 mg/L value and relay #3 (output) is enabled for DO range low 2 when
value is 5.5 mg/L. An aerator is wired to relay #3. e aerator wired to relay #3
energizes until the 5200A receives a DO sensor reading ≥ set point value.
Notes:
For normal operation, the 5200A must be at the Run Screen. Alarm -
Conguring the 5200A