Reset System
Menus → System → Service→ Reset To Factory→ Reset System
Specic systems or all systems listed in the Reset System menu can be reset to fac-
tory defaults.
Use1.
▲
and
▼
to scroll and highlight submenus.
Press2.
to select.
When prompted, “Are you sure?”, select one of the following to conrm, 3.
Press4. to save conguration and return to Run Screen.
Reset Sensors
Menus → System → Service→ Reset To Factory→ Reset Sensors
Specic sensors, aux system, or all sensor and aux systems listed in the Reset Sensors
menu can be reset to factory default.
Use1.
▲
and
▼
to scroll and highlight submenus.
Press2.
to select.
When prompted, “Are you sure?”, select one of the following to conrm, 3.
Conguring the 5200A
Press4. to save conguration and return to Run Screen.
Note:
Perform a sensor calibration aer resetting a sensor to factory default. -
Reset Timers
Menus → System → Service→ Reset To Factory→ Reset Timers
A specic timer, feed timer, or all timers listed in the Reset Timers menu can be
reset to factory defaults.
Use1.
▲
and
▼
to scroll and highlight submenus.
Press2.
to select.
When prompted, “Are you sure?”, select one of the following to conrm, 3.
Press4. to save conguration and return to Run Screen.
Relay Test
Menus → System → Service→ Relay Test
Use the Relay Test menu to verify peripheral devices are wired correctly. Entering
this menu turns o any active relays and disables any sensor, aux, and timer system
relays from energizing regardless of value or condition. Systems are not reset when
the Relay Test menu is exited. To reset systems aer exiting the Relay Test menu,
power cycle the 5200A. Example - an energized conductivity low control relay
(normal mode) is de-energized when entering the Relay Test menu. If the conduc-
tivity value is in control low range when the Relay Test menu is exited, the relay
will not re-energize until the system has a change of condition. e conductivity
system must reach set point, then fall in low control range again before energizing
the control low relay. It is strongly recommended to turn the power o, then on, at
the 5200A aer exiting the Relay Test menu.
Conguring the 5200A