User Manual

YSI 5200A
120
YSI 5200A
121
Reset Clean Probes
Menus → System → Service→ Reset Clean Probes
Enable the clean probe timer, congure the duration, and reset clean probe days
remaining from Reset Clean Probe menu. Selecting reset at this menu resets the
clean probe timer to the user congured value.
icon displays at Run Screen
when the clean probe timer expires.
Reset Clean Probe timer at the menu or the Reset Clean Probes menu.
When prompted “Are you sure?”, select “Yes” to reset the clean probe timer to the
user congured number of days.
Reset To Factory
Menus → System → Service→ Reset To Factory
e Reset to Factory menu resets some or all of the 5200A systems to factory defaults.
For example, if the Clean Probe Timer is user congured to 60 days, performing a
factory reset will reset the Clean Probe Timer to 90 days which is the factory default
setting and not the user congured value. See Appendix B Menu Maps - starting on
page 194 for the factory default settings for all systems.
Conguring the 5200A
Notes:
“Reset All” and “Reset All Systemsresets time format to 12 hour and -
date format to MMDDYYYY.
It is strongly recommended that user conguration be maintained in a -
log by the user.
Instrument, aux, sensor, and sux names are not reset at any reset to -
factory menus. Relay names are reset at “reset relaysand reset all
menus.
Perform sensor calibration aer selecting “Reset All” and/or “Reset Sen- -
s o r s”.
Control icons and control menu messages are displayed for 30 seconds -
aer pressing
from reset to factory menus for any system that
was active (energized relay) prior to the factory reset.
Reset All
Menus → System → Service→ Reset To Factory→ Reset All
Resets all systems, sensors, and timers to factory defaults.
Use1.
and
to scroll and highlight submenus.
Press2.
to select.
When prompted, “Are you sure?”, select one of the following to conrm, 3.
Press4. to save conguration and return to Run Screen.
Notes:
Display contrast, backlight timer, and buzzer can only be reset to factory -
defaults through this menu.
Sensors will need calibrated aer performing a factory Reset All and/or
-
Reset Sensor.