119
Disabling the Power Fail System will disable AC power events and disable the General
Alarm in the event of low battery for DC power. Disable Power Fail AC Monitor
when powering 5200A by DC power source.
Enable Low Battery Monitor to trigger the General Alarm in the event DC voltage is
≤ to low battery trigger. e Low Battery Alarm Hold O time is ve (5) seconds and
cannot be user congured. Once the low battery general alarm has been acknowl-
edged by pressing one of 4 sokeys, it is not reset. See General Alarm - page 113.
Verify AC and DC power supplies are properly wired and powered - 1. page
42.
Enter Power Fail Alarm menu and submenus 2.
Menus → System → Power Fail
Use3.
▲
and
▼
to scroll and highlight submenus.
Press4.
to select.
Enable and congure submenus.5.
Press6. to save conguration and return to Run Screen.
Notes:
Press
- to view AC power fail events.
- icon at the Run Screen indicates that the 5200A DC power is ≥ Low
Battery Trigger value with Low Battery Monitor is enabled.
Conguring the 5200A
- icon at Run Screen indicates DC voltage ≤ the Low Battery Trigger
when Low Battery Monitor is enabled.
Low battery alarm is not reset when front panel sokey is pressed. -
AC power fail does not generate a General Alarm. -
Serial Number
Menus → System → Serial Number
e serial number is displayed at the serial number menu. e serial number is
required when contacting YSI customer service. e serial number is also neces-
sary to congure AquaManager. 5200A serial numbers begin with “53-”followed
by seven digits. Serial numbers are generated at the factory and cannot be changed
by the user.
Service
Menus → System → Service
e Service menu includes submenus to congure and Reset Clean Probe Timer,
Reset Systems to factory defaults, and test Relay(s).
Conguring the 5200A