User Manual
YSI 5200A
110
YSI 5200A
111
Congure PC com port number (1-8) in Firmware Updater program.5.
Note:
When using USB-Serial Adapter, verify that the port number assigned -
during driver installation is the same port congured when using Aqua-
Manager Firmware Updater.
Access the YSI web site Downloads page at www.ysi.com to get the latest 6.
version of the 5200A rmware code.
Locate the YSI 5200A rmware and save the le to a temporary directory 7.
on your computer.
Within the lower le box of the Firmware Updater Utility window, browse 8.
to locate the rmware le on local PC and then import it into the utility.
Click OK at the import complete conrmation window.9.
Select “Program Device”. e following Firmware Updater message ap-10.
pears.
Conguring the 5200A
Select “Yes” and the following screens will appear. 11.
On instrument:
On PC:
e 5200A will reboot (power cycle) when rmware upload is complete.12.
Select “ok” when this window appears 13.
Verify version information to conrm downloaded code:14.
Menus → System → Version
If applicable, disable Force Serial Port conguration - 15. page97.
Notes:
During upload, 5200A systems are placed into Service Hold which de- -
energizes any active systems including energized relays. See SVR Hold
Conguring the 5200A