User Manual
YSI 5200A
106
YSI 5200A
107
Press7. to return to Run Screen or to System menu.
Notes:
- Logs cannot be deleted from the 5200A instrument using AquaMan-
ager.
Calibration and Conguration -
logs cannot be cleared from 5200A.
Date/Time
Menus → System → Date/Time
It is important to congure the correct date and time in the 5200A. Four 10-event
daily timers and one feed timer run based on the 5200A date and time. Logs record
events based on the 5200A clock. Date and time display formats are user dened.
Enter Date/Time menu1.
Menus → System → Date/Time
Verify 5200A current date and time.2.
If necessary, use 3.
▲
and
▼
to scroll and highlight submenus.
Press4.
to select.
Congure submenus.5.
Press6. to save setting and return to Run Screen or press
to System menu.
Notes:
Changes to Date and Time menus take place immediately and are saved -
prior to
.
Congured date and time formats are carried throughout menus and -
logs.
Any active (with energized relay) timer system, including 10-event daily -
timers, feed timers, and/or dosing timers, are NOT recalculated when
the 5200A time and/or date is changed. Once a timer relay has been
energized, it runs for the duration and does not reset based on changes
Conguring the 5200A
to the 5200A date and time. Example - timer 1, event 1 is congured to
come on at 9:00am for 10 minutes. Timer relay energizes at 9:00am. At
9:05am the 5200A time is changed to 10:05am (for daylight savings ad-
justment). e timer relay will de-energize aer ve additional minutes.
e 5200A time will be 10:10am. e event log will reect the congured
5200A time when the event occurred. In this example, 9:00am for the
start and 10:10am for the end.
e time must be manually changed for daylight savings time adjustments. -
Daylight savings note: Sensor data log records are not overwritten when
the clock is changed unless the instrument is power cycled aer changing
the data and time. Example - the data log interval is set to 15 minutes.
Current instrument time is 6:00 am. e clock is adjusted to 5:00 am for
daylight savings time. e 5200A sensor log contains four 15 minute in-
terval records from 5:00 am to 6:00 am from prior to time change. Sensor
data logs are not overwritten with new values for the repeat of 5:00 am to
6:00 am 15 minute intervals unless the instrument is power cycled.
Menu congurations are saved to the system by pressing
- to return
to the Run Screen and not when the change is made in the menu. Excep-
tion - Date and time and clear data logs are updated immediately to the
system, not when
is pressed.
Display Settings
Menus → System → Display Settings
Sensor and Aux conditions are displayed at the Run Screen. Use the Display Set-
tings menu to congure how many systems display on one page at the Run Screen.
Up to six systems can be displayed on one Run Screen page. Multiple Run Screen
pages scroll at user congured speed (1-65535 seconds).
Conguring the 5200A