User Manual

YSI 5200A
102
YSI 5200A
103
Go to Sensor Setup Menus  Sensor Setup Aux 1-2 - 6. page 133 to enable
and congure set points, control, alarm, relays, and hold o time.
Press
7. to save conguration.
Note:
e alarm low value must be ≥ the min value. e alarm high value must -
be the max value.
Data Logging
Menus → System →Data Logging
e 5200A saves information in the calibration, sensor, and event logs. Press
at the Run Screen to view the 25 most recent records for each log. e 5200A sensor
log can store 1000 records. e calibration log can store 500 records. e event log
can store 1250 records. Once a 5200A log is full, it will begin overwriting the oldest
records with new records (data wrapping).
Notes:
- Use AquaManager to store and analyze 5200A logs. To avoid gaps in data
use polling intervals based on the user dened data logging conguration.
Set the data log interval so that analyzing the data is meaningful.
- AquaManager users can view Configuration events in the
Conguration log. Conguration
events are created when certain 5200A
conguration changes are made. e conguration log is not accessible at
the 5200A. See AquaManager Online Help for additional information.
Data Logging
Menus → System →Data Logging→Data Logging
e 5200A sensor log stores records for sensors and aux systems. Sensor logs are
time and date stamped based on the user congured logging interval (1-255 min-
utes/hours).
Conguring the 5200A
Enter Data Logging menu1.
Menus → System →Data Logging→Data Logging
Use2.
and
to scroll and highlight submenus.
Press3.
to select.
Enable and congure submenus.4.
Press5. to save conguration and return to Run Screen.
Press to view the 25 most recent sensor logs.
Notes:
Temperature values are stored in °C regardless of display format.
-
Aux digital closed” is logged as Aux 0.00. Aux digital openis logged
-
as Aux 1.00.
e 5200A sensor log records 0.00 resolution for aux analog inputs regard- -
less of the selected display format.
DO values are stored in mg/L & %Sat regardless of selected display
-
format.
Sensor logs record the default system name and not the user congured
-
system name.
Changes to the data logging interval take eect aer the expiration of the -
interval being changed. Example - if the interval is changed from one
hour to one minute, one minute logging will begin aer the expiration of
the current one hour interval. Power cycle the 5200A to make a change
to the data logging interval occur immediately.
Menu congurations are saved to the system by pressing
- to
return to the Run Screen and not when the change is made in the menu.
Exception - Date and time and clear data logs are updated immediately
to the system, not when
is pressed.
e sensor log contains values for all sensor and aux systems regardless -
if the system is disabled.
5200A sensor log values for disabled systems and systems without sen- -
sors are not valid.
e Conductivity sensor log is reported in Siemens (S). 1 S = 1,000 -
millisiemen(mS) = 1,000,000 microsimen (µS). e conductivity sensor
log resolution is 0.00. For example, 68.37 mS is stored as 0.07 S in the
sensor log and 5960 µS is stored as 0.00 S.
AquaManager data log displays 5200A conductivity - values in µS to 0.00
resolution.
Conguring the 5200A