User Manual
YSI 5200A
100
YSI 5200A
101
Analog 0-1VDC and 0-5VDC and 4-20 mA
Menus → System → Congure Aux →Aux 1-2→Input Type →0-1V
→0-5V
→4/20mA
Select the analog input type based on the signal being sent from the device.
Notes:
It is normal to hear an on-board relay “click” when switching to or from
-
the 4-20mA aux input type.
Analog systems will not monitor, control, or alarm properly if a device
-
that operates outside the congured analog range is installed and con-
gured.
Sufx
Menus → System → Congure Aux →Aux 1-2→Sux
Enter the Sux (up to 4 characters) to describe the input type or units. e sux
is displayed at the Run Screen. e sux is typically used to describe the input
signal type.
Example - when using a total dissolved gas pressure sensor, “mmHg” could be used
as a sux for measuring TGP (total dissolved gas pressure) - gure 4.6, page 101.
Min and Max values
Menus → System → Congure Aux →Aux 1-2 →Min
→Max
Min and Max values are user dened. e range is -10000 to 10000. Min and Max
values are used for scaling the input value received from an analog aux input. e
min and max values may be congured to correspond to the minimum input level
(i.e. 0VDC or 4mA) and maximum input level (i.e. 1VDC, 5VDC, or 20mA). e
Min value must be less than Max value.
Example - when using a total dissolved gas pressure sensor (0-5VDC input type)
enter -400 for min value. e 5400 will display “-400” when the input value is 0VDC.
Set the max value to correspond to the maximum input value as +400. e 5200A
will display “400” when input value is 5VDC.
Notes:
e 5200A will monitor and data log based on the “min” and “max” - volt-
age or current signal.
To enter negative values, type the whole number then select “-” on the -
numeric keypad.
Conguring the 5200A
Whenever min and/or max values are changed, go to - Menus → System
→ Sensor Setup → Aux 1-6 → Control Low and Control High menu to
verify control/alarm values are correct for the application.
Display Format
Menus → System → Congure Aux →Aux 1-2→Display Format
Congure the Display format (0, 0.0, or 0.00) based on the resolution to be displayed
on the Run Screen.
Example - when using a total dissolved gas pressure sensor congure, the display
format as 0.0 - gure 4.6, on this page.
In the below screen shot, the Run Screen is displaying aux name “TGP,” sux
“mmHg,” with display format “0.0”. e display value represents an input of 1.875
volts with 0-5VDC analog aux conguration. e min value = “-400”. e max
value = “+400”. e control/alarm low value is “-50”. Alarms are enabled for this
aux input.
gure 4.6
Notes:
Values are displayed based on the display format resolution. Example - if -
the display format is 0.0 and the min value congured is 1.88, then 1.9
will be the value displayed in the min value menu.
5200A aux sensor logs record 0.00 resolution regardless of the selected -
resolution in the aux analog display format conguration.
Wire aux analog devices - 1. page 49.
Enter Aux menu2.
Menus → System → Congure Aux →Aux 1-2
Use3.
▲
and
▼
to scroll and highlight submenus.
Press4.
to select.
Press
5. to save conguration.
Conguring the 5200A