User Manual

YSI 5200A
98
YSI 5200A
99
Non-Networked (Stand Alone) 5200As
Non-networked (stand alone) 5200As must have the termination relay enabled. In
addition, non-networked (stand alone) 5200As must be congured as a master in
the System Communication menu with no slave nodes enabled.
Congure AUX
Menus → System → Congure Aux 1-2
e 5200A has two auxiliary (Aux) inputs. Aux inputs can be congured as digital or
analog (0-1VDC, 0-5VDC, or 4-20 mA) inputs. Congure sux, minimum (Min),
maximum (Max), and display format menus for the analog inputs. Go to Sensor
Setup Menus Sensor Setup Aux 1-2 - page 133 to enable Aux channel and congure
control/alarm settings. Both digital and analog directions are on the following pages.
Submenu functionality is provided aer the conguration directions.
Probe Name
Menus → System → Congure Aux →Aux 1-2→Probe Name
Use the alpha/numeric keypad to name aux inputs. Congured name(s) appear
throughout menus and at Run Screen - gure 4.6, page 101.
Note:
Aux names can also be congured at Menus  System  Name Devices/ -
Sensors.
Input Type
Menus → System → Congure Aux →Aux 1-2→Input Type
Select the input type for the device(s) wired to aux system(s). Aux Sensor Setup
(page 133) menu selections are based on the aux input type congured at the Sys-
tem menu.
Conguring the 5200A
Digital
Menus → System → Congure Aux →Aux 1-2→Input Type→Digital
Use the digital input when wiring an external device to control or alarm when the
contacts of the external switch are open or close. A common application would be
to detect high or low water levels using a oat switch.
Notes:
Sux, Min, Max, and Display Format are not available for digital in- -
puts.
In the 5200A sensor log, aux digital closed” is logged as Aux 0.00. Aux -
digital “open” is logged as Aux 1.00.
Wire aux digital devices - 1. page 49.
Enter Aux menu2.
Menus → System → Congure Aux →Aux 1-2
Use3.
and
to scroll and highlight submenus.
Press4.
to select.
Congure digital input type and name.5.
Press6. to save conguration and return to Run Screen.
Go to Sensor Setup Menus  Sensor Setup Aux 1-2 - 7. page 133 to enable
and congure control/alarm on condition, control/alarm relay, and hold
o time.
Press
8. to save conguration.
Conguring the 5200A