User Manual
YSI 5200A
98
YSI 5200A
99
Non-Networked (Stand Alone) 5200As
Non-networked (stand alone) 5200As must have the termination relay enabled. In
addition, non-networked (stand alone) 5200As must be congured as a master in
the System Communication menu with no slave nodes enabled.
Congure AUX
Menus → System → Congure Aux 1-2
e 5200A has two auxiliary (Aux) inputs. Aux inputs can be congured as digital or
analog (0-1VDC, 0-5VDC, or 4-20 mA) inputs. Congure sux, minimum (Min),
maximum (Max), and display format menus for the analog inputs. Go to Sensor
Setup Menus Sensor Setup Aux 1-2 - page 133 to enable Aux channel and congure
control/alarm settings. Both digital and analog directions are on the following pages.
Submenu functionality is provided aer the conguration directions.
Probe Name
Menus → System → Congure Aux →Aux 1-2→Probe Name
Use the alpha/numeric keypad to name aux inputs. Congured name(s) appear
throughout menus and at Run Screen - gure 4.6, page 101.
Note:
Aux names can also be congured at Menus System Name Devices/ -
Sensors.
Input Type
Menus → System → Congure Aux →Aux 1-2→Input Type
Select the input type for the device(s) wired to aux system(s). Aux Sensor Setup
(page 133) menu selections are based on the aux input type congured at the Sys-
tem menu.
Conguring the 5200A
Digital
Menus → System → Congure Aux →Aux 1-2→Input Type→Digital
Use the digital input when wiring an external device to control or alarm when the
contacts of the external switch are open or close. A common application would be
to detect high or low water levels using a oat switch.
Notes:
Sux, Min, Max, and Display Format are not available for digital in- -
puts.
In the 5200A sensor log, aux digital “closed” is logged as Aux 0.00. Aux -
digital “open” is logged as Aux 1.00.
Wire aux digital devices - 1. page 49.
Enter Aux menu2.
Menus → System → Congure Aux →Aux 1-2
Use3.
▲
and
▼
to scroll and highlight submenus.
Press4.
to select.
Congure digital input type and name.5.
Press6. to save conguration and return to Run Screen.
Go to Sensor Setup Menus Sensor Setup Aux 1-2 - 7. page 133 to enable
and congure control/alarm on condition, control/alarm relay, and hold
o time.
Press
8. to save conguration.
Conguring the 5200A