User Manual
YSI 5200A
96
YSI 5200A
97
Send Test Email Msg
Menus → System → Communications → Send Test Email Msg
Send a test email to verify proper conguration of ethernet menus. Test emails are
only transmitted from congured master 5200As.
Note:
Exit to Run Screen is required before the test email is sent. -
Example of a 5200A test email sent from an ethernet device:
Subject Line: 02/04/10 03:50:05 PM YSI5200A Email Alarm Test Message
From: myname@xxx-xxxx.com
Sent: Wed, Feb 04, 2010 3:49 pm
To: myname@xxx-xxxx.com
CC: sweet49@xxxx.com
Email Retry Delay
Menus → System → Communications → Email Retry Delay
Congure the Email Retry Delay for the time interval that emails should be retrans-
mitted aer the initial alarm email is sent. e initial alarm email is sent immediately
aer the general alarm hold o time or aer the aux hold o time. Emails will be
retransmitted until one of the bottom front panel sokeys is pressed, regardless if
the alarm condition still exists.
Note:
On a network, emails are retransmitted according to the slave retry delay -
conguration of the slave in alarm and not the master’s email retry delay
conguration. Please note, a slave Email Retry Delay menu will display
(Not available) because of its slave conguration; however, this feature
is available and has a default value of 15 minutes that can be congured
by the user.
e Retry Delay should be congured at intervals ≥ 15 minutes. -
AquaManager alarm emails are retransmitted based on the 5200A slave(s) -
and master email retry delay conguration.
Baud Rate RS232
Menus → System → Communications → Baud Rate
Congure the Baud rate (for RS232 connections only) in the 5200A to match the
baud rate congured in AquaManager. Use the default rate of 115200 for fastest
communication speed.
Conguring the 5200A
Enter Network menu1.
Menus → System → Communications → Network →Baud Rate
Press2.
▲
or
▼
to scroll and highlight a selection.
Press3.
to conrm selection.
Press4. to save conguration and return to Run Screen.
Force Serial Port
Menus → System → Communications → Force Serial Port
Enable the Force Serial Port when the 5200A is communicating via RS232 (direct)
to a PC and when the 5200A has a serial to ethernet device.
Wire RS232 connection - 1. page 58
Enter Network menu2.
Menus → System → Communications → Force Serial Port
Use the
3. sokeys to congure o or on.
Press4. to save conguration and return to Run Screen.
Notes:
RS232 connection is required to upload rmware using the Firmware -
Updater.
e RS232 icon
- blinks at the Run Screen when in Force Serial Port
mode.
Conguring the 5200A