When DHCP is enabled, the serial to ethernet device acquires SMTP and -
DNS addresses during initialization.
- AquaManager requires a static IP address; therefore, DHCP must be
disabled when using AquaManager. Consult the network administrator
to obtain a static IP address.
Power cycle the 5200A whenever DHCP is enabled. -
IP Address
Menus → System → Communications → Ethernet → IP Address
e IP address can only be congured when DHCP is disabled.
Notes:
- AquaManager users should program the Gateway address either by force
serial port in the AquaManager Communication menu or locally at the
5200A. Poll node using RS232 communication to upload conguration
information including: static IP, port, subnet, and gateway addresses.
When DHCP is enabled, the IP address is not updated at the IP address -
menu.
Power cycle the 5200A whenever a new IP address is congured. -
Subnet Mask
Menus → System → Communications → Ethernet → Subnet Mask
A Subnet mask address can only be congured when DHCP is disabled. e subnet
mask number helps to dene the relationship between the host (computers, rout-
ers, switches, etc.) and the rest of the network. Consult the network administrator
when conguring this menu.
Notes:
When DHCP is enabled, the Subnet Mask address is not updated at the -
Subnet Mask menu.
- AquaManager users should program the Subnet mask either by force serial
port in the AquaManager Communication menu or locally at the 5200A.
Poll node using RS232 communication to upload conguration informa-
tion including: static IP, port, subnet, and gateway addresses.
Gateway
Menus → System → Communications → Ethernet → Gateway
e Gateway address can only be congured when DHCP is disabled. A gateway IP
refers to a device on a network which sends local network trac to other networks.
Consult the network administrator when conguring this menu.
Conguring the 5200A
Notes:
When DHCP is enabled, the Gateway address is not updated at the -
Gateway menu.
AquaManager users should program the Gateway address either by force -
serial port in the AquaManager Communication menu or locally at the
5200A. Poll node using RS232 communication to upload conguration
information including: static IP, port, subnet, and gateway addresses.
Port
Menus → System → Communications → Ethernet → Port
e default port number for the ethernet module is 10001. Consult the network
administrator when conguring this menu.
Note:
- AquaManager users should program the port number either by force
serial port in the AquaManager Communication menu or locally at the
5200A. Poll node using RS232 communication to upload conguration
information including: static IP, port, subnet, and gateway addresses.
DNS Address
Menus → System → Communications → Ethernet → DNS
Domain Name Service (DNS) numbers are IP addresses that a workstation or server
uses to refer to specic servers that resolve domain names to IP addresses.
Notes:
When DHCP is enabled, the serial to ethernet device will acquire SMTP -
and DNS addresses during initialization.
When DHCP is enabled, the DNS address is not updated at the IP ad- -
dress menu.
Email Addresses
Menus → System → Communications → Ethernet → Email Address
→ Email Address
→ Email Address
→ CC/Email Address
Up to four email addresses can be congured for 5200A emails. At a minimum,
the rst email address and the CC/email address must be congured. If the CC/
email address is not congured, 5200A alarm emails will not contain any alarm
information. e email from’ address will be the email address entered in the rst
Email Address menu.
Conguring the 5200A