User Manual
YSI 5200A
82
YSI 5200A
83
Aer one minute, verify that the pH and temperature values show no 9.
signicant change for approximately 30 seconds. en, press to
continue the calibration OR press to escape calibration and return
to calibration menu without saving.
10. Using the arrow keys to select numbers on the numeric keypad,
enter the pH buer value. Press
to conrm each number selection.
Press 11. to continue OR press to escape the calibration and
return to calibration menu without saving.
For one-point calibration skip to step12. 18.
Rinse the calibration cup and sensors, then ll the calibration cup with 13.
the 2nd buer solution for a 2-point calibration.
Replace probe in calibration cup and let settle for at least 30 seconds.14.
Aer one minute, verify the pH and temperature values show no signicant 15.
change for approximately 30 seconds. en, press to continue the
calibration OR press to escape calibration and return to calibration
menu without saving any of the calibration.
16. Enter the value of the 2nd buer solution. Press
to conrm
each number selection.
Press
17. to continue OR press to escape the calibration and
return to calibration menu without saving.
Press
18. to save calibration and return to Run Screen.
Remove calibration storage cup from probe assembly.19.
Rinse the sensor in tap or puried water before continuing to calibrate 20.
other sensors or reinstalling the probe guard.
Conguring the 5200A
System Menu
Menus → System
e System Menu contains 17 submenus. Use System submenus to congure non-
sensor and non-timer systems. System submenus are shown below. Press
at
the Run Screen to enter the Main Menu. Use ▲ ▼ to scroll and highlight a System
submenu. Press
8 to select and enter a highlighted menu. Conguration changes
are saved when
is pressed.
Notes:
See - Appendix 3 “Menu Maps” for system defaults and ranges (where ap-
plicable) - starting on page 194.
Menu congurations are saved to the instrument only aer
- is
pressed and not when the change is made in the menu. If is not
pressed, the changes made to the conguration will not be saved. Excep-
tion - Date and time and clear data logs are updated immediately to the
system; however, YSI recommends pressing
in this instance as
well in order to save any other changes that may have been made to the
conguration.
Conguring the 5200A