User Manual

YSI 5200A
78
YSI 5200A
79
Follow these instructions to create a 100% water-saturated air environment 6.
for DO sensor calibration:
For the 5562 probe assembly:
Place approximately 3 mm (~ 1/8”) of water in the bottom of the •
calibration/storage cup.
Next, engage only 1 or 2 threads of the calibration/storage cup to •
ensure the DO sensor is vented to the atmosphere. Note: Make
sure that the DO sensor is not immersed in water and that there
are no water droplets on the membrane.
For the 5561 probe assembly:
Moisten the sponge in the clear calibration/storage chamber with •
a small amount of clean water. Do not use too much water! e
idea is to create a 100% humid environment.
With the sensor guard installed, place the sensor into the clear •
calibration/storage chamber. Make sure that the DO sensor is
not immersed in water and that there are no water droplets on
the membrane.
Enter DO calibration menu7.
Menus → Calibration→ DO→Sat
Allow approximately 10 minutes for the air in the calibration cup to become 8.
completely water saturated and for the temperature to equilibrate.
Aer 10 minutes, verify that the DO reading shows no signicant change 9.
for approximately 30 seconds. en, press to complete the calibra-
tion OR press
to abort the calibration without saving.
Press 10. to save calibration and return to Run Screen.
Remove the calibration/storage cup from probe assembly.11.
Rinse sensor in tap or puried water and place probe assembly in monitor-12.
ing location or continue calibrating other sensors.
Conguring the 5200A
Directions to Calibrate DO Sensor mg/L Value
A large bucket can be used when calibrating the DO mg/L reading. Be sure to
provide sample ow across the membrane by physically stirring the probe or using
a stirrer such as a stir bar. e DO sensor requires a sample ow rate of at least 3
inches (~7.62 cm) per second if using the 2.0 mil PE membrane (blue) and at least
6 inches (~15.24 cm) per second if using the 2.0 mil Teon membrane (gray).
Connect probe assembly -
1. page 60
Read 2. Getting Ready to Calibrate - page 73
Wait at least 10-15 minutes aer the 5200A is powered on, or aer the 3.
DO sensor is serviced, before calibrating. (is allows enough time for
the DO sensor to stabilize).
If not using a conductivity sensor or if the conductivity sensor is disabled, 4.
enter the salinity (ppt) value of the water to be monitored in DO Sensor
Setup menu. is will result in accurate DO mg/L readings.
Menus → Sensor Setup→ DO→ppt
Place the DO probe in a water sample with a known DO concentration 5.
(the DO mg/L value can be determined from another recently calibrated
DO instrument or from a Winkler titration).
Enter the DO calibration menu.6.
Menus → Calibration→ DO→mg/L
Let probe stabilize for approximately one minute.7.
Aer one minute, verify that the DO value shows no signicant change for 8.
approximately 30 seconds. en, press to complete the calibration
OR press to cancel the calibration without saving.
9. Using the arrow keys to select numbers on the numeric keypad,
enter the known DO value in mg/L. Press
to conrm each number
selection.
Press
10. to conrm the entered value OR press to abort the
calibration without saving any changes.
Conguring the 5200A