User Manual
YSI 5200A
58
YSI 5200A
59
Notes:
Firmware can only be uploaded to an instrument using a local RS232 con- -
nection. See Downloader - page 109.
On a network, install and congure RS232 or the serial to ethernet device -
(hardware and wiring) on the master node only.
Email Alarming
Email alarming can be congured in one of two ways:
- from a PC running the AutoPoll Program that is included with AquaMan-
ager soware (PC must have internet connection). is is referred to as
AquaManager email.
via a serial to ethernet* device with TCP/IP connection at master 5200A. -
is is referred to as 5200A email.
*Serial to ethernet connection requires access to a SMTP server accessible by the
local network.
RS232 Communication
Use Category 5e patch cable from RS232 port at location H - gure 3.29 of I/O
board to PC using RJ45 to DB9 adapter.
Directions to Wire Direct RS232 PC Connection
WARNING: Disconnect external power to the unit before wiring.
AVERTISSEMENT : déconnectez l’alimentation externe de l’unité avant d’eectuer
un câblage quelconque.
Perform steps 1 - 4 of 5200A installation.1.
Open front panel - 2. page 39.
Feed direct connect cable through drilled hole of rubber grommet and 3.
to location H.
Run high and low voltage cables through separate bulkhead and con-4.
duit.
Connect one end of communication cable to RJ45 at location 5. H. Connect
other end to PC com port using RJ45 to DB9 adapter.
Installation and Wiring
Do not connect or disconnect IDC ribbon cable
when 5200A is powered. Serious damage can occur
.
Figure 3.29
Complete other wiring to 5200A.
6.
Close front panel - 7. page 40.
Test RS232 communication.8.
Notes:
- When using a USB-Serial Adapter, verify that the port number assigned
during driver installation is the same port congured when using Aqua-
Manager and/or the Firmware Updater.
When making RS232 cable runs greater than 100 feet (30.5 m), use an RS232 -
Line Extender/Booster. Mount the booster in an area where it is protected
from moisture.
Proper PC com port number must be congured when using AquaManager -
or Firmware Updater programs.
Test RS232 Communications
Apply power to 5200A. Insert node into AquaManager program. Poll node. If
wired correctly, AquaManager will get 5200A conguration and logs.
Ethernet (TCP/IP) Module Communication
e optional serial to ethernet device must be installed on the 5200A control board
when using TCP/IP for communication. Complete installation instructions are
provided with the optional serial to ethernet device.
Installation and Wiring