Firmware can only be uploaded to an instrument using a local RS232 con- -
nection. See Downloader - page 109.
On a network, install and congure RS232 or the serial to ethernet device -
(hardware and wiring) on the master node only.
Email Alarming
Email alarming can be congured in one of two ways:
- from a PC running the AutoPoll Program that is included with AquaMan-
ager soware (PC must have internet connection). is is referred to as
AquaManager email.
via a serial to ethernet* device with TCP/IP connection at master 5200A. -
is is referred to as 5200A email.
*Serial to ethernet connection requires access to a SMTP server accessible by the
local network.
RS232 Communication
Use Category 5e patch cable from RS232 port at location H - gure 3.29 of I/O
board to PC using RJ45 to DB9 adapter.
Directions to Wire Direct RS232 PC Connection
WARNING: Disconnect external power to the unit before wiring.
AVERTISSEMENT : déconnectez l’alimentation externe de l’unité avant d’eectuer
un câblage quelconque.
Perform steps 1 - 4 of 5200A installation.1.
Open front panel - 2. page 39.
Feed direct connect cable through drilled hole of rubber grommet and 3.
to location H.
Run high and low voltage cables through separate bulkhead and con-4.
duit.
Connect one end of communication cable to RJ45 at location 5. H. Connect
other end to PC com port using RJ45 to DB9 adapter.
Installation and Wiring
󱂈
󱂈
󱂈
󱂈
󱂉
󱂊
󱂋
󱂌
󱂍
󱂎
󱂏
󱂐
󱂑
󱂒
󱂓
Do not connect or disconnect IDC ribbon cable
when 5200A is powered. Serious damage can occur
.
Figure 3.29
Complete other wiring to 5200A.
6.
Close front panel - 7. page 40.
Test RS232 communication.8.
Notes:
- When using a USB-Serial Adapter, verify that the port number assigned
during driver installation is the same port congured when using Aqua-
Manager and/or the Firmware Updater.
When making RS232 cable runs greater than 100 feet (30.5 m), use an RS232 -
Line Extender/Booster. Mount the booster in an area where it is protected
from moisture.
Proper PC com port number must be congured when using AquaManager -
or Firmware Updater programs.
Test RS232 Communications
Apply power to 5200A. Insert node into AquaManager program. Poll node. If
wired correctly, AquaManager will get 5200A conguration and logs.
Ethernet (TCP/IP) Module Communication
e optional serial to ethernet device must be installed on the 5200A control board
when using TCP/IP for communication. Complete installation instructions are
provided with the optional serial to ethernet device.
Installation and Wiring