User Manual

YSI 5200A
44
YSI 5200A
45
󱂈
󱂈
󱂈
󱂈
󱂉
󱂊
󱂋
󱂌
󱂍
󱂎
󱂏
󱂐
󱂑
󱂒
󱂓
Do not connect or disconnect IDC ribbon cable
when 5200A is powered. Serious damage can occur
.
Figure 3.17
Complete other wiring to 5200A.5.
Close front panel - 6. page 40.
Test AC power.7.
Test AC Power
Turn on main power. e enabled sensor values will display on the Run Screen aer
the initial YSI splash screen. If applicable, install and congure DC power backup
page 45. Verify AC power at the Menus → System →Version menu - page 125.
See Power Fail Menu - page 117 for AC power event log information.
Wire DC Power and backup power
e 5200A may be powered by most 9.0-16.5 VDC source that provides 800mA of
current and is isolated from mains supply by double or reinforced insulation. e
DC power source is user-supplied and can come from a variety of choices including
lead acid, gel cell, or UPS (with 12VDC transformer) external batteries. DC power
wires are terminated at location B on the I/O Board - gure 3.18.
Notes:
e 5200A does not charge batteries. Quality assurance maintenance -
procedures should be established if batteries are to be fully powered when
used as back-up power source.
Power supply voltage above 16.5 VDC may permanently damage the -
5200A.
Installation and Wiring
When back up voltage falls below 9.0 volts, the 5200A ceases to operate -
properly.
Some switching DC power supplies not supplied by YSI may result in -
noisy readings.
Directions to Wire DC Power/DC Backup Power
WARNING: Disconnect external power to the unit before wiring. Follow
all manufacturer’s safety and installation instructions when providing power to
the 5200A via a DC power supply.
AVERTISSEMENT : déconnectez lalimentation externe de lunité avant
d’eectuer le câblage. Suivez toutes les consignes de sécurité et d’installation du
fabricant lorsque vous fournissez de l’alimentation au modèle 5200A par un bloc
d’alimentation à courant continu.
Perform steps 1 - 3 of 5200A installation.1.
Open front panel - 2. page 39.
Feed DC power cable through drilled hole in rubber grommet to location 3.
B. The terminal strip is removable.
CAUTION: Run high and low voltage cables through separate bulkhead and
conduit.
Terminate DC power wires to pins (-) and (+) according to manufacturer 4.
instructions and any applicable local electrical codes.
󱂈
󱂈
󱂈
󱂈
󱂉
󱂊
󱂋
󱂌
󱂍
󱂎
󱂏
󱂐
󱂑
󱂒
󱂓
Do not connect or disconnect IDC ribbon cable
when 5200A is powered. Serious damage can occur
.
Figure 3.18
Installation and Wiring