User Manual

YSI 5200A
38
YSI 5200A
39
Figure 3.13
Accessing the I/O Board
e I/O board is accessible when the 5200A front panel is removed. e letters on
the metal plate attached to the I/O board identify pluggable terminal strips on the
I/O board. Letters on the metal plate correspond to the wiring diagram located
on the inside of the front panel - gure 3.14. A ribbon cable connects the display
board (mounted in the front panel) to the I/O board. When the 5200A front panel
is removed, a lanyard protects against stress on the ribbon cable. See Open Front
Panel to Access I/O Board - page 39.
Installation and Wiring
Figure 3.14
Open Front Panel to Access I/O Board
WARNING: Disconnect external power to the unit before doing any wir-
ing.
AVERTISSEMENT : connectez l’alimentation externe de l’unité avant
douvrir le panneau avant.
Loosen the 4 mounting screws connecting the front panel to the back 1.
enclosure. Captive screws remain in front panel aer loosening.
Carefully li the 5200A front panel from the back enclosure. Front panel 2.
should be placed to the le of enclosure. Allow the lanyard to hold the
weight of the front panel. e rubber gasket seated in the front panel
should remain in the front panel lid.
Installation and Wiring