condition. Timer systems are not reset. To reset timer systems, turn power o and
then on at the 5200A. Press
to exit Service Hold.
e Run Screen displays a clock count down of the Service Hold remaining time.
Notes:
- If a timer is active (relay energized) at service hold, the relay will not reen-
ergize aer service hold expires regardless if timer time has not expired.
ere is only one “on” time for all timer systems.
- Entering Downloader mode places 5200A into Service Hold.
- “Inst. Service Hold” is displayed on any AquaManager map icon associ-
ated with an instrument in Service Hold. AquaManager alarm emails
are not sent for any alarm condition for a 5200A that is in Service Hold.
- 5200A alarm emails are not sent during service hold.