Owner's manual
YSI 5200A
169
YSI 5200A
168
A Reset Clean Probe message will appear at the Msg’s Menu if the clean probe timer
expires. Selecting “Yes” at “Are you sure?” resets the clean probe timer to the user
congured number of days.
Note:
- User congured sensor and relay names do not appear in log menus.
Calibration Logs
Calibration logs store sensor calibration data. Calibration logs are date and time
stamped and contain sensor and calibration information for completed calibra-
tions.
Certain calibration data is not available at the 5200A. Use AquaManager to view
and save calibration data to the PC. In addition to sensor system, time, date, and
calibration method, AquaManager calibration logs include user input 1, user input
2, and the original and new slope/oset information when applicable.
Event Logs
Control, Alarm, and Timer relays assigned to 5200A systems create events when
they energize and de-energize. In addition, AC power fail and ethernet failure to
initialize errors create events. ese events are recorded in the Event log. Events
identify the 5200A system, are time stamped, and record relay on and o times.
e General Alarm does not create an event or control message unless a General
Alarm relay is enabled.
Notes:
- Enable event logging in the System menu. See Event Logging - page 104
for additional information.
Sensor Logs
Enabled systems record sensor data based on data logging menu conguration.
Sensor logs identify sensor system, are date and time stamped, and record sensor
value at user congured intervals.
Notes:
- Data logging must be enabled in the System Menu. See Data Logging -
page 102 for additional information.
Svr Hold Menu
When servicing sensors or doing tank maintenance, it may be necessary to shut
o all systems (including relays) because sensor values will not be accurate. Press
at Run Screen to enter Service Hold menu. When in Service hold, all relays
are de-energized until the Service Hold time expires. e Service Hold time is user
selectable. Enable and congure service hold time at Menus → System→Service
Hold - page 124. Aer service hold time expires or is cancelled, sensor and aux
systems are reset causing relays to energize/de-energize based on the system value/
Conguring the 5200A Conguring the 5200A