Owner's manual

YSI 5200A
167
YSI 5200A
166
Notes:
- Always manually poll the 5200A aer uploading conguration changes
and verify that the uploaded information is retained. When conguration
changes are uploaded using AquaManager at the same time conguration
changes are being made by a user in the menus at the 5200A, AquaMan-
ager uploads may not be saved depending on the timing of keystrokes
at the 5200A.
- e Run Screen must be displayed during normal operation. If the
5200A is not at the run screen during AquaManager uploading, press-
ing
at the 5200A will only save certain conguration changes.
See Chapter 5 AquaManager - page 172 and AquaManager Online Help for ad-
ditional information.
Control Menu
Press at the Run Screen to display a list of energized sensor and aux 5200A
system relays. e Control menu identies the active sensor or aux system and
control and/or alarm condition. See Sensor Setup Controls and Alarms - page 132
for additional information.
Notes:
- Press
to display time stamped detail of control and alarm relay(s).
- Alarm message appears at the Control menu for aux and sensor systems
when alarms are enabled regardless if an alarm relay is assigned.
- Active timer (10 event timer and feed timer) systems are not displayed
at control menu.
- General alarm systems are not displayed at the Control menu regardless
if general alarm relay is enabled.
- Use to scroll list when scroll bar appears at menu.
- Conductivity and pH Control Range Low message remains at the
Conguring the 5200A
menu when the conductivity or pH system is disabled and a control low
relay is energized. Power cycle the 5200A to clear the control low range
message at the Msg’s menu.
When no 5200A control or alarm system(s) are active, the below screen is dis-
played.
Msg’s Menu
Calibration, Sensor, and Event logs contain 5200A system information. e 5200A
Sensor log stores 1000 records. e Calibration log stores 500 records. e Event
log stores 1250 records. Once the log memory is full, the oldest records are over-
written with more recent records. e number shown in ( ) at the Msgs menu
reects the total number of records in the log. e 25 most recent records for each
log are displayed at the
menu. Use to scroll and press
to view
time stamped detail record.
Use AquaManager to capture all records from 5200A logs and store them to a PC
database. Polling times should be determined by logging intervals. Example - if the
sensor log is recording values every minute and the sensor log holds 1000 records,
polling should occur at least every 27 hours. is would allow for all records to be
saved to the PC before being overwritten on the 5200A. Set the data log interval so
that analyzing data based on the congured interval will be helpful. Poll at intervals
that dont result in gaps in data.
Conguring the 5200A