Owner's manual
YSI 5200A
137
YSI 5200A
136
alarm icons, hear audible alarm and/or receive alarm notications sent via email or
cell phone (SMS) - page 95. Congure high and low alarm values to set alarm
thresholds. Alarm system(s) becomes active when value is ≤ low alarm value or ≥
high alarm value. Enabled alarm relays energize if alarm condition exists for the
general alarm hold o time. Alarm relays are energized until one of the four so-
keys
is pressed regardless of sensor value. Pressing
one of four sokeys resets the alarm system. Alarm systems become active again
if, aer the alarm hold o period, the alarm condition still exists. Alarm hold o is
congured in the General Alarm menu.
When the General Alarm is enabled, (with or without enabled relay), the general
alarm system becomes active when any system alarm is active. See General Alarm
- page 117.
System low alarm range starts at -1 signicant digit below low control value. High
alarm range starts at +1 signicant digit above high control value.
Notes:
- It is important to keep the 5200A at the Run Screen during normal
operation. Alarm functionality (email alarms, activation of enabled
alarm relays, alarm icons, and buzzer) is suspended when 5X00 is not
at Run Screen.
- Alarm icon(s) display at the Run Screen when alarm condition(s) exists
for longer than the general alarm hold o time. See Icons - page 70.
- Press
at Run Screen to display active sensor and aux system relays.
-
to display events created by control and alarm relays - see Event
Log - page 107.
- Alarm relays are not energized if the 5200A sensor value reads “Over”
or “Under.”
- See Wire Relays - page 46 for information on wiring alarm output
devices.
Validation Error and Autox - Set Point Control Only
When an invalid value is entered at the numeric keypad a Validation Error window
displays. An invalid value is either a value outside the range of the sensor system or
a value that breaches other settings within the sensor system. Autox occurs when
a high or low control value is entered that breaches the set point value. Examples
are provided below:
Low range control value must be < set point value. High range control value must
be > set point value. Low alarm value must be < low control value. High alarm
value must be > high control value.
Conguring the 5200A
Example - entering “63” for DO set point results in validation error because the
valid range for the DO sensor system is 0.00 to 60.0 mg/l. User must congure
value within valid range. In addition, the value cannot breach respective control
and alarm system values.
Example - conguring set point value of 8.00 mg/L with range low value of 8.00
mg/l results in validation error. Valid set point range is displayed at validation error
display. Press
to return to numeric keypad to enter valid value.
Autox occurs when a high or low control value is entered that breaches the set point
value. Autox changes the high or low control value by one least signicant digit
away from the set point value. e set point value is never adjusted.
Example - Entering 8.00 mg/L as the range low value when the set point value is
8.00 mg/L results in an auto x of the control value. e invalid control value is
displayed at the sensor setup menu. Pressing
is not an option aer saving
a value at numeric keypad. Press
from the sensor setup menu to continue
with the autox. A validation error window is displayed to show that the autox has
occurred. Press
to return to sensor setup menu. Verify that the autoxed
values are congured for user’s specic application and recongure as needed. Press
to save conguration and return to Run Screen.
Conguring the 5200A