AquaManager and the System Menu
AquaManager users can congure most 5200A system menus from a PC. Aer poll-
ing the 5200A, use “Systems Management” from the node window to view and upload
5200A congurations. System menus are congured at the Setup and Input tabs.
Conguring the 5200A
Notes:
- Always manually poll the 5200A aer uploading conguration changes
and verify the uploaded information is retained. When conguration
changes are uploaded using AquaManager at the same time congura-
tion changes are being made by a user in the menus on the 5200A, the
AquaManager uploads may not be saved depending on the timing of
keystrokes at the 5200A.
- e Run Screen must be displayed during normal operation. If the
5200A is not at the run screen during AquaManager uploading, pressing
at the 5200A will only save certain conguration changes.
See Chapter 5 AquaManager - page 172 and AquaManager online help for ad-
ditional information.
Sensor Setup
Menus → Sensor Setup→Aux 1
→Aux 2
→Conductivity
→DO
→ORP
→pH
→Temperature
→Salinity
Aux, Conductivity, DO, ORP, pH, and temperature system set points, control mode
(set point or PID/PWM), alarm and sensor system relays are congured in the
Sensor Setup menu. In addition, the salinity system is enabled in this menu. Four
onboard relays can be congured for control and alarm output devices. Enabled
relays energize when a control or alarm system is active. Alarm notications can
be sent via email or cell phone (SMS messaging) - page 95.