Reset Clean Probe timer at the menu or the Reset Clean Probes menu.
When prompted “Are you sure?”, select “Yes” to reset the clean probe timer to the
user congured number of days.
Reset To Factory
Menus → System → Service→ Reset To Factory
e Reset to Factory menu resets some, or all, of the 5200A systems to factory
defaults. For example, if the Clean Probe Timer is user congured to 60 days,
performing a factory reset will reset the Clean Probe Timer to 90 days which is the
factory default setting and not the user congured value. See Appendix B Menu
Maps - starting on page 201 for factory default settings for all systems.
Conguring the 5200A
Notes:
- “Reset All” and “Reset All Systems” resets time format to 12 hour and
date format to MMDDYYYY.
- It is strongly recommended that user conguration be maintained in a
log by the user.
- Instrument, aux, sensor, and sux names are not reset at any reset to
factory menus. Relay names are reset at “reset relays” and “reset all
menus.
- Perform sensor calibration aer selecting “Reset All” and/or “Reset Sen-
s o r s”.
- Control icons and control menu messages are displayed for 30 seconds
aer pressing
from reset to factory menus for any system that
was active (energized relay) prior to the factory reset.
Reset All
Menus → System → Service→ Reset To Factory→ Reset All
Resets all systems, sensors, and timers to factory defaults.
1.
Use
and
to scroll and highlight submenus.
2.
Press
to select.
3. When prompted, “Are you sure?”, select one of the following to conrm,
4. Press to save conguration and return to Run Screen.
Notes:
- Display contrast, backlight timer, and buzzer can only be reset to factory
defaults through this menu.
- Sensors must be recalibrated aer performing a factory Reset All and/
or Reset Sensor.
Reset System
Menus → System → Service→ Reset To Factory→ Reset System
Specic systems or all systems listed in the Reset System menu can be reset to fac-
tory defaults.
Conguring the 5200A