Owner's manual
YSI 5200A
125
YSI 5200A
124
Reset Clean Probe timer at the menu or the Reset Clean Probes menu.
When prompted “Are you sure?”, select “Yes” to reset the clean probe timer to the
user congured number of days.
Reset To Factory
Menus → System → Service→ Reset To Factory
e Reset to Factory menu resets some, or all, of the 5200A systems to factory
defaults. For example, if the Clean Probe Timer is user congured to 60 days,
performing a factory reset will reset the Clean Probe Timer to 90 days which is the
factory default setting and not the user congured value. See Appendix B Menu
Maps - starting on page 201 for factory default settings for all systems.
Conguring the 5200A
Notes:
- “Reset All” and “Reset All Systems” resets time format to 12 hour and
date format to MMDDYYYY.
- It is strongly recommended that user conguration be maintained in a
log by the user.
- Instrument, aux, sensor, and sux names are not reset at any reset to
factory menus. Relay names are reset at “reset relays” and “reset all”
menus.
- Perform sensor calibration aer selecting “Reset All” and/or “Reset Sen-
s o r s”.
- Control icons and control menu messages are displayed for 30 seconds
aer pressing
from reset to factory menus for any system that
was active (energized relay) prior to the factory reset.
Reset All
Menus → System → Service→ Reset To Factory→ Reset All
Resets all systems, sensors, and timers to factory defaults.
1.
Use
▲
and
▼
to scroll and highlight submenus.
2.
Press
to select.
3. When prompted, “Are you sure?”, select one of the following to conrm,
4. Press to save conguration and return to Run Screen.
Notes:
- Display contrast, backlight timer, and buzzer can only be reset to factory
defaults through this menu.
- Sensors must be recalibrated aer performing a factory Reset All and/
or Reset Sensor.
Reset System
Menus → System → Service→ Reset To Factory→ Reset System
Specic systems or all systems listed in the Reset System menu can be reset to fac-
tory defaults.
Conguring the 5200A